Various principles are used for measuring the weight of objects. The following briefly describes the operating principles and features of the two most popular methods, "electromagnetic type" and "load cell type."

    ● Electromagnetic Type

    This is also called the "electromagnetic balance method." With mechanical balances, the sample is placed at one end of the beam and the weight is placed at the other end, and the value of the weight when both are perfectly balanced becomes the mass of the sample. With electromagnetic type balances, an electrical force (electromagnetic force) is applied instead of a placed weight to balance the beam. The amount of electricity required for balancing the beam changes according to the weight of the sample that is placed. The amount of current when the beam is perfectly balanced is detected, and the mass is obtained from that detected value.

    ● Load Cell Type (electrical resistance wire method)

    One end of an object (elastic body) made of aluminum and shaped as shown in the figure is fixed in place, and the sample is placed on the other end. The weight of the sample causes the elastic body to flex. The amount of flex causes the strain gauges attached to the elastic body to expand and contract, changing the amount of electricity that is output (strictly speaking, the resistance value). The mass is then obtained from that amount of electricity.

    ● Comparison Between Electromagnetic Type and Load Cell Type

     
    Electromagnetic Type
    Load Cell Type
    Advantages
    • ・High accuracy
    • ・Simple structure
    • ・Even large models are easy to make
    Disadvantages
    • ・Complex structure
    • ・Difficult to downsize
    • ・Accuracy is limited
    Applications
    • ・Ultra-precision balances such as analytical balances
    • ・Small, cheap balances that require only moderate accuracy
    • ・Large balances
    Name of Series
    • ・Microbalances and analytical balances (e.g. AW/AX/AY series)
    • ・Top-loading precision balances (e.g. BX/BW, BL series)
    • ・EL series, LDS series, AHW height/weight scales, etc.

    Movies

    Related applications