ASTM D8421-22 Standard Test Method for Determination of Per- and Polyfluoroalkyl Substances (PFAS) in Aqueous Matrices by Co-solvation Followed by Analysis Using the Shimadzu LCMS-8060NX

Download

User Benefits

- The LCMS-8060NX easily meets and exceeds method performance criteria of ASTM D8421 for 44 PFAS and 24 surrogates. - Optimized chromatography and MS conditions for excellent peak shape for improved precision and accuracy. - ASTM D8421 is a simple extraction procedure validated by ASTM for the analysis of PFAS in wastewater samples.

Introduction

ASTM International published ASTM D84211 for the analysis of 44 per- and polyfluorinated alkyl substances and 24 la- beled isotopes in non-potable water samples. This method extracts the substances in a 1+1 ratio of sample and methanol, filters and then measures the targeted com- pounds using external standard calibration liquid chro- matography-tandem mass spectrometry (LC/MS/MS). The minimum reporting limit is 10 ng/L with an analytical range of 10 - 400 ng/L for most compounds. The method requires standard solutions be prepared by the laboratory from neat compounds. To save a laboratory’s time and effort from preparing stock standards individually, we optimized the method using com- mercially available native and labeled calibration standard mixes. Additionally, we optimized chromatography, achiev- ing better peak shape for early-eluting compounds, such as PFBA and PFPrA. This application news summarizes the performance of the Shimadzu LCMS-8060NX Liquid Chromatography Mass Spec- trometer (LC/MS/MS) for all analytes listed in ASTM D8421. Results meet or exceed the requirements outlined in the method. The reporting limit (RL) for the test method is defined as an integer value that is equal to the concentration of the lowest calibration standard.

14 de enero de 2024 GMT

Related Products

Some products may be updated to newer models

Related Solutions