Language
Español
English
0

    W-AD Series Analytical Balance

    (Equipped Standard with Smart Automatic Doors and Touchless Sensors)

    Model AP225W-AD AP135W-AD AP225WD-AD AP125WD-AD AP324W-AD AP224W-AD
    Capacity 220g 135g 220g / 102g 120g / 52g 320g 220g
     Minimum Display  0.01mg 0.1mg / 0.01mg 0.1mg
    Calibration Weight Built-in (*1)
    External Calibration Weight Range for Span Calibration (Recommended weight value)
    95 to 220.00090g
    (200g)
    45 to 135.00090g
    (100g)
    95 to 220.00090g
    (200g)
    45 to 120.00090g
    (100g)
    95 to 320.0090g
    (300g)
    95 to 220.0090g 
    (200g)
     Repeatability (at weighing capacity) 
    (*2)
    0.015mg (up  to 20g)
    0.03mg (up to 100g)
    0.05mg (up to 200g)
    0.05mg 0.1mg / 0.05mg 0.1mg / 0.02mg 0.15mg 0.1mg
    Repeatability
    (for low loads) (*2, *3)
    0.01mg
    (for 5g load)
    0.1mg
    (for 20g load)
    0.1mg
    (for 10g load)
     Minimum Weight (*2, *3) 20mg (*2) 200mg
     Linearity (*2) ±0.1mg ±0.2mg / ±0.1mg ±0.2mg / ±0.05mg ±0.3mg ±0.2mg
     Response Time for Trace Measurements (*4, *5) 2 sec.
     Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
     USB Host (Type A)
     USB Device (Type B)
    Recipe Compounding
     HPLC Buffer Solution Preparation
     mol Conversion Function
     Sample (Concentration) Preparation
     Inspection Support Function
     Clock-CAL
    Automatic Doors
    Touchless Sensors
    Adjustable Windbreak Plate -
    Ionizer
    Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
     Sensitivity Stability Against Temperature Range ±2ppm / ℃ (10~30℃)
    Pan Size Approx. 91 mm dia.
    Body Dimensions Approx. W215 × D411 × H346 mm (power supply unit included) Approx. W215 × D367 × H346 mm
    Weight Approx. 9.7kg Approx. 8.6kg
    Display OEL display (dot matrix)
    Rated Electric Power Supply DC12V 1.5A
    Power Supply Input (AC Adapter) AC100-240V 480mA 50/60Hz (*7)
    Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

    *1 Minimum display 0.01 mg models provide two internal weights as standard.

    *2 Measurement conditions of W-AD series (0.01 mg models only) are as follows.
     - Set the adjustable windbreak plate in the lowest position
     - With a shield plate configured around the pan

    *3 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
       The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

    *4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

    *5 The response time value is typical.

    *6 Non-condensing.

    *7 Depending on the attached AC adapter.

    W Series Analytical Balance

    Model AP225W AP135W AP125WD AP225WD AP124W AP224W AP324W
    Capacity 220g 135g 120g / 52g 220g / 102g 120g 220g 320g
    Minimum Display 0.01mg 0.1mg / 0.01mg 0.1mg
    Calibration Weight Built-in (*1)
    External Calibration Weight Range for Span Calibration (Recommended weight value)
    95 to 220.00090g
    (200g)
    45 to 135.00090g
    (100g)
    45 to 120.00090g
    (100g)
    95 to 220.00090g
    (200g)
    45 to 120.0090g
    (100g)
    95 to 220.0090g
    (200g)
    95 to 320.0090g
    (300g)
     Repeatability (at weighing capacity)
    (*3)
    0.015mg (up  to 20g)
    0.03mg (up to 100g)
    0.05mg (up to 200g)
    0.05mg 0.1mg / 0.02mg 0.1mg / 0.05mg 0.1mg 0.15mg
    Repeatability
    (for low loads) (*2, *3)
    0.01mg (for 5 g load) 0.1mg
    (for 5g load)
    0.1mg
    (for 10g load)
    0.1mg
    (for 20g load)
     Minimum Weight (*2, *3) 20mg 200mg
    Linearity (*3) ±0.1mg ±0.2mg / ±0.05mg ±0.2mg / ±0.1mg ±0.2mg ±0.3mg
     Response Time for Trace Measurements (*4, *5) 2 sec.
     Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
     USB Host (Type A)
    USB Device (Type B)
    Recipe Compounding
    HPLC Buffer Solution Preparation
    mol Conversion Function
    Sample (Concentration) Preparation
    Inspection Support Function
    Clock-CAL
    Internal Windbreak Plate ○ (optional)
    Ionizer (optional)
    Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
    Sensitivity Stability Against Temperature Range ±2ppm/℃ (10~30℃)
    Pan Size Approx. 91 mm dia.
    Body Dimensions Approx. W213 × D411 × H345 mm (power supply unit included) Approx. W213 × D367 × H345 mm
    Weight Approx. 7.9 kg Approx. 7.0 kg
    Display OEL display (dot matrix)
    Rated Electric Power Supply DC12V 1.0A
    Power Supply Input (AC Adapter) AC100 - 240 V / 320mA / 50/60Hz
    Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

    *1 Minimum display 0.01 mg models provide two internal weights as standard.

    *2 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
       The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

    *3 The value is the result of a test with the internal windbreak plate.

    *4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

    *5 The response time value is typical.

    *6 Non-condensing.

    *7 Depending on the attached AC adapter.

    X Series / Y Series Analytical Balance

    Series X Series Y Series
    Model AP324X AP224X AP124X AP324Y AP224Y AP124Y
    Capacity 320g 220g 120g 320g 220g 120g
    Minimum Display 0.1mg
    Calibration Weight Built-in None
    External Calibration Weight Range for Span Calibration (Recommended weight value)
    95 to 320.0090g
    (300g)
    95 to 220.0090g
    (200g)
    45 to 120.0090g
    (100g)
    95 to 320.0090g
    (300g)
    95 to 220.0090g
    (200g)
    45 to 120.0090g
    (100g)
    Repeatability (at weighing capacity) 0.15mg 0.1mg 0.15mg 0.1mg
    Repeatability
    (for low loads) (*1)
    0.1mg (for 20g load) 0.1mg (for 10g load) 0.1mg (for 5g load) 0.1mg (for 20g load) 0.1mg (for 10g load) 0.1mg (for 5g load)
    Minimum Weight (*1) 200mg
    Linearity ±0.3mg ±0.2mg ±0.3mg ±0.2mg
     Response Time for Trace Measurements (*2, *3) 2 sec.
     Response Time (*3) 2 sec.
     USB Host (Type A)
    USB Device (Type B)
    Recipe Compounding
    HPLC Buffer Solution Preparation
    mol Conversion Function
    Sample (Concentration) Preparation
    Inspection Support Function
    Clock-CAL
    Ionizer
    Operational Temperature / Humidity Range 5~40℃ 20 to 85% (*4)
    Sensitivity Stability Against Temperature Range ±2ppm/℃ (10 to 30℃)
    Pan Size Approx. 91 mm dia.
    Body Dimensions Approx. W213 × D367 × H345 mm
    Weight Approx. 7.0 kg Approx. 6.5 kg
    Display OEL display (dot matrix)
    Rated Electric Power Supply DC12V 1.0A
    Power Supply Input (AC Adapter) AC100 - 240 V / 320mA / 50/60Hz (*5)
    Input / Output Terminal RS-232C (D-Sub 9P plug)  USB device (Type B)  Ionizer RS-232C (D-Sub 9P plug)  USB device (Type B) 

    *1 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
       The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

    *2 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

    *3 The response time value is typical.

    *4 Non-condensing.

    *5 Depending on the attached AC adapter.