    TOC-4200

    Measurement items

    TC, NPOC
    Optional: IC, POC, TOC (= TC-IC, = NPOC + POC), TN

    Data storage function

    Store 20,000 past on-line measured values (equivalent to one year's data at 30-minute measuring cycle).
    Offers trend graph displays.

    Dimensions

    W550 x D383 x H1240 mm (excluding protrusions)
    (W22 x D15 x H49 inches)

    Construction

    Indoor wall mounting (optional stand available)

    Ambient temperature

    Within 1 to 40℃

    Power supply

    AC 100 to 240 V ±10%, 10 A, 50/60 Hz

    Carrier gas

    Pressurized air, high-purity air, oxygen*4; supply pressure: 250 to 300 kPa
    Optional: High-purity nitrogen*4

    Communication function

    Compatible with RS-485, RS-232C (option), and Modbus
    Use the optional Web monitoring unit set to browse measured values and alarms on a Web browser over a LAN.

    Contact input

    Start calibration, stop measurement, start on-line measurement, reset alarm, measurement start flow line 1 to 6, halt measurement (output contact capacity: 1.4 V DC min., 10 mA min.)

    Contact output signal

    Alarm signals
    Major failure, instrument alarm, power cut, CPU alarm, concentration alarm (upper limit/lower limit/upper upper limit/lower lower limit), measurement halted
    Event signals
    Maintenance, measurement ready, on-line operation, measuring sample, calibrating, regenerating catalyst, measuring control sample, interrupt sample measurement, halted, analog output trigger, analog output flow line recognition signal, measured flow line recognition signal, sampling pump control output (contact capacity: 2 A, 30 V DC / 2 A, 250 V AC; non-inductive load)

    Analog output

    Selectable between 4 – 20 mA DC or 0 – 16 mA DC (isolated outputs)
    With attached option: Up to 12 outputs (load resistance: 500Ω max.)

    Printer

    Thermal printer, 40 characters per line; 110 mm chart width (optional)

    Recorder

    Recording width 100 mm, 6-dot recorder (optional)

    Data storage device

    Store measured values or measurement conditions to USB memory*3.
    (USB 1.1 or 2.0 USB memory; FAT16 or FAT32; no encryption)

    EPA TOC removal rate calculation function

    TOC removal rate calculation according to the United States EPA regulations (Part IV 40 CFR Part 9, 141 and 142, 1998) (when a 2- or more-stream switching option is provided)

    Measurement principle

    680℃ combustion catalytic oxidation - NDIR detection method

    Load calculation function

    Using the analog input board (option) permits load calculations. (Non-isolated analog inputs; load resistance: 100 ohms)

    Conversion function

    Conversion with a linear expression using required indices (Conversion expression must be determined in advance.)

    Display, operation

    Color LCD with touch screen

    Stream switching function

    Streams 1 to 6 can be switched (with the relevant option)

    Automatic calibration function

    Automatic calibration using standard solution (Dilution water is used as the zero standard solution for zero-calibration.)
    Options can be attached to use up to eight standard solutions and to permit automatic calibration with up to a 5-point multi-point calibration curve.

    Sample dilution function

    Dilution within syringe; dilution factor 2 to 50
    Dilution accuracy: Within ±2 % (x2 to x20); within ±5 % (x21 to x50)

    IC removal method

    Acidify/sparge within syringe

    Sample injection method

    Syringe pump/sliding injection port

    Measurement cycle

    4 minutes min. (NPOC, residual IC 2 % max.)*2

    Zero stability

    Within ±2 % f.s. /day (temperature variation within 5℃)*1

    Span stability

    Within ±2 % f.s. /day (temperature variation within 5℃)*1

    Repeatability

    Within ±2 % f.s.*1

    Measurement range

    From 0 - 5 to 0 - 1,000 mgC/L f.s. (0 to 20,000 mgC/L f.s. with dilution function)

    Weight

    Approx. 70 kg (154 lbs.)

    *1) With automatic settings
    *2) When measured sample is potassium hydrogen phthalate solution and operations to prepare for the next measurement are performed.
    *3) If TN option is installed, oxygen and high-purity nitorogen cannot be used.

    Read more
    Show Less