    NOA-7100 - Specs

    Transportable NOx-O2 Gas Analyzer

    Standard type

    Two standard types are available

    Type 1

    Measurement components

    NOx, O2

    Wireless data transmission

    Equipped

    Sample gas conditions (*5)

    [ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa;
    Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less

    External drain separator

    Fitted

    Weight (analyzer)

    Approx. 16 kg

    Dimensions

    W260 x H452 x D420 mm (does not include protrusions)

    Power Requirements

    Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz,
    100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)

    Permissible ambient humidity (*4).

    Relative humidity 90%RH or less

    Permissible ambient temperature (*4)

    5-40℃; must not be exposed to direct sunlight or radiant heat

    Data output to external media (*3)

    Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection.

    Transmission Output

    0-1VDC, 3 channels, insulated output (individual channels not insulated)

    Measurement range

    NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/4000 ppm
    O2: 0 - 5/ 10/ 25 vol%

    Display

    Liquid crystal display, able to display 3 items at a time

    Gas sample amount

    Approx. 2 L/min

    Response time (*2)

    NOx: Approx. 20/30/60 sec (Switching is configurable.)
    O2: Approx. 30/60 sec (Switching is configurable.)

    Linearity

    Within ±2% of full scale

    Span Drift

    Within ±1.5% of full scale per day

    Zero Drift

    Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)

    Repeatability

    Within ±0.5% of full scale

    Measurement principle

    NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method

    External Signal Input

    0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)

    Read more
    Show Less

    Type 2

    Measurement components

    NOx, NO, NO2 (*1)

    Wireless data transmission

    Equipped

    Sample gas conditions (*5)

    [ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa;
    Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less

    External drain separator

    Not fitted

    Weight (analyzer)

    Approx. 16 kg

    Dimensions

    W260 x H452 x D420 mm (does not include protrusions)

    Power Requirements

    Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz,
    100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)

    Permissible ambient humidity (*4).

    Relative humidity 90%RH or less

    Permissible ambient temperature (*4)

    5-40℃; must not be exposed to direct sunlight or radiant heat

    Data output to external media (*3)

    Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection.

    Transmission Output

    0-1VDC, 3 channels, insulated output (individual channels not insulated)

    Measurement range

    NOx, NO, NO2: 0 - 100/ 250/ 500/ 1000/ 2500/ 4000 ppm

    Display

    Liquid crystal display, able to display 3 items at a time

    Gas sample amount

    100 mL/min

    Response time (*2)

    NOx/NO: Within 90 sec

    Linearity

    Within ±2% of full scale

    Span Drift

    Within ±1.5% of full scale per day

    Zero Drift

    Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)

    Repeatability

    Within ±0.5% of full scale

    Measurement principle

    NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method

    External Signal Input

    0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)

    Read more
    Show Less

    Special specifications

    In addition to the standard model, it is possible to specify constituents, and if the small flow rate set, communication set, or NOx-NO switching solenoid valve set is provided.

    Measurement
    components
    1 constituent (NOx) or 2 constituents (NOx, O2) (*5)
    O2 cannot be measured if small flow rate set is selected. See small flow rate set page.
    NO and NO2 can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.
    See NOx-NO switching solenoid valve set page.
    Measurement
    range
    NOx: 0-25/50/100/250/500/1000/2500/4000 ppm
    O2: 0-5/10/25vol%
    NOx measurement range differs if small flow rate set is selected.
    See small flow rate set page.

    Small flow rate set : Not fitted

    Response time (*2)

    NOx: Approx. 20/30/60 sec (Switching is configurable.)
    O2: Approx. 30/60 sec (Switching is configurable.)

    Gas sample amount

    Approx. 2 L/min

    Repeatability

    Within ±0.5% of full scale

    Span Drift

    Within ±1.5% of full scale

    Measurement components

    NOx or NOx/O2
    NO and NO2 (*1) can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.

    Measurement range

    NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/ 4000 ppm
    O2: 0-5/ 10/ 25 vol%

    Read more
    Show Less

    Small flow rate set : Fitted

    Response time (*2)

    NOx/NO: Within 90 sec

    Gas sample amount

    100 mL/min

    Repeatability

    Within ±0.5% of full scale

    Span Drift

    Within ±1.5% of full scale

    Measurement components

    NOx
    NO and NO2 (*1) can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.

    Measurement range

    NOx: 0-100/ 250/ 500/ 1000/ 2500/ 4000 ppm

    Read more
    Show Less

    Communication set : Not fitted

    Wireless data transmission

    Not equipped

    Data output to external media (*3)

    Not equipped

    Communication set : Fitted

    Wireless data transmission

    Equipped

    Data output to external media (*3)

    Measurement data stored in the internal memory of the analyzer can be output to USB flash drives in CSV file format; can also be output to computer via Wi-Fi or LAN cable connection.

    NOx-NO switching solenoid valve set : Not fitted

    Measurement components

    NOx

    NOx-NO switching solenoid valve set : Fitted

    Measurement components

    NOx, NO, NO2 (*1)

    This special accessory switches the flow path inside the analyzer between a path that flows through the converter (NOx flow path) and one that does not flow through the converter (NO path). The switch cycle is selectable from auto (configurable from 1 to 99 minutes), manual, and fixed. If there is significant variation over time in concentrations of NOx, NO, or NO2 relative to the switching cycle, NO2 concentration measurement accuracy diminishes.

    Basic specifications are as follows.

    Basic specifications

    Measurement principle NOx: Atmospheric pressure chemiluminescence method
    O2: Limiting current type zirconia method
    Repeatability Within ±0.5% of full scale
    Zero drift Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)
    Span drift Within ±1.5% of full scale
    Linearity Within ±2% of full scale

    The following specifications are the same as for the standard model.
    Display, transmission output, permissible ambient temperature, permissible ambient humidity, power requirements, dimensions, weight (body), sample gas conditions, external signal input.

    *1:

    NO2 is measured by calculation (NOx-NO).

    *2:

    Time that indicates 90% response from gas flowing from sample gas inlet (flowing gas switching time + electrical system response time). With Type 2, this will be the response when NOx-NO switching is fixed.

    *3:

    Check the Shimadzu website for smart devices and USB flash drives for which operation has been confirmed.

    *4:

    Permissible ambient temperature and humidity are indicated by the shaded part of the figure.

    *5:

    Sample gas conditions and coexisting gases are summarized below. Please contact your Shimadzu representative if the conditions are not satisfied.