Language
Español
English
0

    EDX-8100 - Specs

    Energy Dispersive X-ray Fluorescence Spectrometer

    EDX-7000/8000/8100

    Measurement principle

    X-ray fluorescence spectrometry

    Measurement method

    Energy dispersion

    Target samples

    Solids, liquids, powders

    Measuring range

    6C to 92U

    Sample size

    W 300 x D 275 x approx.H 100 mm (excluding radiuses)

    Maximum sample mass

    5kg (200g per sample when using turret, Gross mass 2.4kg)

    Dose rate

    1 µSv/h or less.

    Read more
    Show Less

    X-ray generator

    X-ray tube

    Rh target (Standard model/Premium model)*1

    Voltage

    4 kV to 50 kV

    Current

    1 µA to 1000 µA

    Cooling method

    Air-cooled (with fan)

    Irradiated area

    Automatic switching in four stages: 1, 3, 5, and 10 mm diameter
    Automatic switching in four stages: 0.3, 1, 3, and 10 mm diameter*2

    Primary filters

    Five types (six, including the open position), automatic replacement

    *1 Option for EDX-8100
    *2 Microsoft® Office is not included.

    Read more
    Show Less

    Detector

    Type

    Silicon drift detector (SDD)

    Liquid nitrogen

    Not required (electronic cooling)

    Sample chamber

    Measurement atmosphere

    Air, vacuum*1, helium (He)*2

    Sample replacement*

    12-sample turret

    Sample observations

    Semiconductor camera

    *1 Option for EDX-8100
    *2 Option for EDX-7000

    Data processor

    Memory

    4 GB min.

    HDD

    250 GB min.

    OS

    Windows™ 10 Pro (64-bit)*

    Optical drive

    Super multi drive

    CPU

    Intel Core i5 or above

    * Microsoft Office is not included.

    Read more
    Show Less

    Software

    Qualitative analysis

    Measurement/analysis software

    Quantitative analysis

    Calibration curve method, correction for coexistent elements, FP method,film FP method, background FP method

    Matching software

    Intensity/content

    Utilities

    Automatic calibration functions (energy calibration, FWHM calibration)

    Others

    Instrument status monitoring function,
    analysis results tabulation function

    Read more
    Show Less

    Installation

    Temperature

    10 °C to 30 °C (temperature fluctuation rate 2 °C/hour max.,
    temperature fluctuation range: 10 °C max.)

    Relative humidity

    40 % to 70 % (no condensation)

    Power supply

    100-240 V AC ±10 %, 2 A earthed socket

    Dimensions

    W 460 x D 590 x H 360 mm

    Weight

    Approx. 45 kg

    Options

    Vacuum measurement unit, helium purge unit, turret unit, screening analysis kits

    Read more
    Show Less