    Injection Ports

    Split/Splitless Injection Unit (SPL)

    Enables tool-free column connection. Septum replacement is also possible.

    • Sealing mechanism ClickTek nut
      Injection mode Split, Splitless, High Pressure Injection
      Split ratio Up to 9999.9:1
    • Pressure range 0 to 1035 kPat
      Maximum operating temperature 450 ℃

     
    Supports capillary columns from 50 μm to 530 μm I.D. / Electronic septum purge comes built-in / Gas saver mode reduces gas consumption / Deactivation process (Option)

    Detectors

    Flame Ionization Detector (FID)

    The FlD shows good response for all organic compounds.
    It is characterized by high robustness and a wide linear dynamic range

    • Minimum Detectable
      Quantity (MDQ)
      < 1.5 pg C/s (dodecane)
      Linear dynamic range 1 × 107 (± 10 %)
      Max acquisition rate 2 ms (500 Hz)
      Maximum operating
      temperature
      450 ℃
      Automatic extinguishing
      detection/reignition
      Standard-compatible
    • Flow rate settings Makeup (N2 or He) : 0 to 100 mL/min
      H2 : 0 to 100 mL/min
      Air : 0 to 1000 mL/min

     

    Flame Photometric Detector (FPD)

    The detector of choice for trace-level analysis of phosphorus (P), sulfur (S) and tin (Sn) compounds.

    • Minimum Detectable
      Quantity (MDQ)
      < 55.0 fg P/s (tributyl phosphate)
      Minimum Detectable
      Quantity (MDQ)
      < 2.5 pg S/s (dodecanethiol)
      Easy switching of luminous
      filters
      P, S, Sn
      Dynamic range Tributyl phosphate (P) : 1 × 104
      Dodecanethiol (S) : 1 × 103
      Max acquisition rate 2 ms (500 Hz)
      Max operating temperature 450 ℃
    • Flow rate settings H2 : 0 to 250 mL/min
      Air : 0 to 1000 mL/min

     

    Electron Capture Detector (ECD)

    The detector of choice for trace-level analysis of electrophilic compounds such as halogenated, organometallic and nitro compounds.

    • Minimum Detectable
      Quantity (MDQ)
      < 4.0 fg/s (Lindane)
      Dynamic range 1 × 105
      Max acquisition rate 2 ms (500 Hz)
    • Max operating temperature 400 ℃
      Flow rate settings ECD gas (N2, Ar) : 0 to 200 mL/min

    LabSolutions Series

    GC Software