    DPiMS™ QT - Specs

    Kit for Direct Probe Ionization Mass Spectrometer

    Hardware

    DPiMS QT controller

    Control of the DPiMS QT unit (Installed in the mass spectrometer)

    DPiMS QT unit

    Probe Voltage: ±5 kV max. (set voltage)
    Probe Stroke: 46.30 mm max.
    Number of Extraction Cycles: 300 max.
    Probe Speed: 300 mm/sec max.
    Probe Acceleration: 1 G max.
    Probe Stop Time: 60,000 msec max.
    Probe drive cycle: 0.75 to 3 Hz

    Software

    PESI MS Solution LabSolutions

    Used to select analytical conditions and start analysis
    Used to specify MS analysis method settings and analyze data

    (Probe Control Software)

    Used to specify probe control method settings

    Consumables

    Probe

    Probe (set of 10 or 50 pcs.)

    Plate

    Plates available for both liquid and biological samples (set of 100 pcs. each)

    Installation requirements*

    Room temperature

    18 to 28 °C

    Humidity

    20 to 70 % (No-condensing and no-discharge due to static electricity)

    Others

    Should be installed in an environment without dust, oscillation, electromagnetic noise, corrosive gas and electromagnetic interference.

    *Equivalent to the installation requirement of mass spectrometer.