English
Español
English
0

    Shim-pack Velox LC Columns - Features

    HPLC Column

    Column Chemistries

    Combining highly efficient core shell particle technology with a wide range of surface chemistries provides you with the best opportunity for optimal resolution. With different chemistry characteristics, Shim-pack Velox columns are suitable for use in a wide variety of applications and challenging separations.

    Velox
      SP-C18 C18 Biphenyl PFPP HILIC
    USP Classification L1 L1 L11 L43 L3
    Ligand type Sterically protected C18 C18 Biphenyl Pentafluorophenyl propyl None
    Particle Size (μm) 1.8, 2.7, 5 1.8, 2.7, 5 1.8, 2.7, 5 1.8, 2.7, 5 2.7
    Pore size (Å) 90 90 90 90 90
    Surface Area 1.8 μm 125 m2/g 125 m2/g 125 m2/g 125 m2/g -
    2.7 μm 130 m2/g 130 m2/g 130 m2/g 130 m2/g 130 m2/g
    5 μm 100 m2/g 100 m2/g 100 m2/g 100 m2/g -
    Carbon Load 1.8 μm 7 % 9 % 7 % 4 % -
    2.7 μm 7 % 7 % 7 % 4 % N/A
    5 μm 5 % 5 % 5 % 3 % -
    End-Cap No YES YES No No
    pH range 1.0-8.0 2.0-8.0 1.5-8.0 2.0-8.0 2.0-8.0
    Max Pressure 1.8 μm 100 MPa* 100 MPa* 100 MPa* 100 MPa* -
    2.7 μm 60 MPa 60 MPa 60 MPa 60 MPa 60 MPa
    5 μm 40 MPa 40 MPa 40 MPa 40 MPa -
     * For maximum lifetime, recommended maximum pressure for 1.8 μm particles is 80MPa.​
     

    Column Selection Guide for Different LC Platforms

    Column particle sizes and column volumes affect chromatography results significantly if the column configuration does not match the LC system. As column particle size is reduced, or the column volume (ID and/or the length of the columns) decreases, the necessity for a lower dispersion system is increased.
    Choosing the optimal column configuration for your LC system allows you to achieve improved chromatography. 
    The following table summarizes the starting recommendations of column configuration for each LC system.

     
     

    Prominence

    Prominence

    Prominence-i

    Nexera XR

    Nexera-i

    Nexera X2

    LC System HPLC UHPLC-like UHPLC
    Particle size 2.7 μm & 5 μm 2.7 μm 1.8 μm & 2.7 μm
    Column I.D. 4.6 mm (3.0 mm) 3.0 mm (2.1 mm) 2.1 mm
    Column Length 100-250 mm 50-100 mm ≤150 mm
     

    Lot to Lot Reproducibility

    We understand that lot to lot consistent performance of columns is required to maximize your laboratory performance. 
    From one lot to the next, every Shim-pack Velox column you purchase will perform consistently.