    i-Series - Specs

    High Performance Liquid Chromatograph

    LC-2050C (UV model)

    Pump

    Degassing unit

    Five Lines: Mobile phase 4 + Rinse solution 1 [Volume 400 μL]

    Pumping method

    Parallel-type double plunger

    Pulsation

    ≤ 0.1 MPa [1.0 mL/min,10 MPa, Water]

    Flow rate setting range

    0.0001 to 10 mL/min

    Flow rate accuracy

    ≤ ±1% or ≤ ±2 μL/min, below whichever is greater
    [0.01 to 2 mL/min, Specified condition]
    ≤ ±2%
    [2 to 5 mL/min, Specified condition]

    Flow rate precision

    < 0.06 %RSD or < 0.02 minSD, below whichever is greater

    Configuration

    Four-solvent low-pressure gradient

    Gradient / range of set concentrations

    0 to 100%, in 0.1% steps

    Gradient / concentration accuracy

    ±0.5% [0.1 to 2 mL/min, 1 to 20 MPa, Specified condition]

    Gradient / concentration precision

    ±0.1% [1 mL/min,10 MPa, Specified condition]

    Maximum pressure

    50 MPa [0.0001 to 3 mL/min]
    44 MPa [3.0001 to 5 mL/min]
    22 MPa [5.0001 to 10 mL/min]

    System Delay Volume

    650 μL
    [Option: 460 μL, 1100 μL]

    Read more
    Show Less

    Autosampler

    Sample injection method

    Total-volume sample injection

    Injection volume accuracy

    ±1% [50 μL, N = 10]

    Injection volume setting range

    0.1 to 100 μL
    [Option: 0.1 to 50 μL, 1 to 500 μL, 1 to 2,000 μL]

    Injection volume reproducibility

    RSD < 0.20% [5.0−2000 μL]
    RSD < 0.25% [2.0−4.9 μL]
    RSD < 0.5% [1.0−1.9 μL]
    RSD < 1.0% [0.5−0.9 μL]

    Cross-contamination

    0.0025% [Caffeine, Specified condition]

    Injection cycle time

    Min. 14 sec [Specified condition]

    Samples for processing

    336 [1 mL], 216 [1.5 mL], 112 [4 mL], 4 [MTP/DWP]

    Sample cooler

    4 to 45°C
    [Room temperature needs to be 30°C or lower and the humidity 70% or less to go down to 4°C. Room temperature needs to be 15°C or higher to go up to 45°C.]

    Injection linearity

    > 0.9999
    [1 to 100 μL, Specified condition]

    Read more
    Show Less

    Column Oven

    Heating and cooling method

    Forced air circulation method

    Containable column size

    Six columns 10 cm long
    Three columns 30 cm long
    One 30 cm long column if a CTO unit (optional) for i-Series low-temperature
    analysis is installed. (Two columns up to 15 cm long or one column 30 cm long
    if an FCV tubing kit for i-Series low-temperature analysis is also installed.)

    Temperature control range

    Room temperature −12 °C (−15 °C if an optional unit is installed) to 90 °C

    Temperature control precision

    ±0.1°C

    Temperature accuracy

    ±0.8°C (Specified condition)

    Flow rate switching valve

    Max. 1 pc

    Read more
    Show Less

    UV Detector

    Wavelength range

    190 to 700 nm

    Spectral bandwidth

    8 nm

    Wavelength accuracy

    ≤ ±1 nm

    Wavelength Reproducibility

    ≤ ±0.1 nm

    Noise

    ≤ ±2.5 × 10−6 AU, (250 nm, Specified condition)

    Drift

    ≤ 100 × 10−6 AU/h [250 nm, Specified condition]

    Simultaneous monitoring of 2 wavelengths

    Enable [Any 2 wavelengths of 190 to 370 nm or 371 to 700 nm]

    Linearity

    Up to 2.5 AU (5%)

    Sampling rate

    Up to 100 Hz

    Light source

    Deuterium (D2) lamp

    Flow Cell

    12 μL (10 mm, TC), 12 MPa

    Option cell

    High-Speed: 8 μL [10 mm, TC]
    Semi-micro: 2.5 μL [5mm, TC]

    Read more
    Show Less

    Miscellaneous

    Dimensions

    W410×H605×D500 mm [Not including reservoir tray height]

    Weight

    63 kg

    Available pH range

    1 to 13

    Materials for parts in contact with liquids

    Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby, sapphire, Hastelloy® C, GFP, ceramic, PFA, quartz, PPS

    Read more
    Show Less