0

    Pump

    Degassing unit

    Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 µL)

    Pumping method

    Parallel-type double plunger

    Pulsation

    ≤ 0.1 MPa (1.0 mL/min,10 MPa, Water)

    Flow rate setting range

    0.0001 to 10 mL/min

    Flow rate accuracy

    ≤ ±1% or ±2 µL/min, below whichever is greater (0.01 to 2 mL/min, Specified conditions)
    ≤ ±2% (2 to 5 mL/min, Specified conditions)

    Flow rate precision

    ≤ 0.06 %RSD or ≤ 0.02 minSD, below whichever is greater

    Configuration

    Four-solvent low-pressure gradient

    Gradient / range of set concentrations

    0 to 100%, in 0.1% steps

    Gradient / concentration accuracy

    ±0.5% (0.1 to 2 mL/min, 1 to 20 MPa, Specified conditions)

    Gradient / concentration precision

    ±0.1% (1 mL/min, 10 MPa, Specified conditions)

    Maximum pressure

    Packed column side : 34 MPa (0.0001 to 5 mL/min) , 22 MPa (5.0001 to 10 mL/min)
    Slide-in column side : 7 MPa (0.0001 to 0.3 mL/min)

    System Delay Volume

    approximately 615 µL (Packed column side)
    approximately 620 µL (Slide-in column side) (Option: none)

    Read more
    Show Less

    Autosampler

    Injection method

    Total-volume sample injection

    Injection volume accuracy

    ±1% (50 µL, N=10)

    Injection volume setting range

    0.1 to 100 µL (Option:0.1 to 50 µL)

    Injection volume reproducibility

    RSD ≤0.20% (5.0 to 100 µL), RSD ≤0.25% (2.0 to 4.9 µL)
    RSD ≤0.5% (1.0 to 1.9 µL), RSD ≤1.0% (0.5 to 0.9 µL)

    Cross-contamination

    0.0025% (Caffeine, Specified conditions)

    Injection cycle time

    Min. 14 sec (Specified conditions)

    Samples for processing

    336 (1 mL), 216 (1.5 mL), 112 (4 mL), 4 (MTP/DWP)

    Sample cooler

    4 to 45°C

    Injection linearity

    ≥0.9999 (1 to 100 µL, Specified conditions)

    Read more
    Show Less

    Column Oven

    Heating and cooling method

    Forced air circulation method

    Containable column size

    Slide-in column
    Packed column (1 pieces at 10 to 25 cm)

    Temperature control range

    Room temperature −5 to 60°C,
    Setting range 4 to 60°C (To 50°C for slide-in column)

    Temperature control precision

    ±0.1°C

    Temperature stability

    ±0.8°C (Specified conditions)

    Flow rate switching valve

    Not available

    Read more
    Show Less

    PDA Detector

    Wavelength range

    190 to 800 nm

    Spectral resolution

    1.4 nm (Specified condition)

    Slit width

    1.2 nm, 8 nm

    Device resolution

    0.6 nm/pixel

    Number of photodiode array elements

    1024

    Wavelength accuracy

    ≤ ±1 nm

    Noise level

    ≤ ±3 × 10−6 AU (250 nm, reference: 350 nm, Specified conditions)

    Drift

    ≤ 500 × 10−6 AU/h (250 nm, reference: 350 nm, Specified conditions)

    Linearity

    Up to 2 AU (5%)

    Sampling rate

    Up to 100 Hz

    Light source

    Deuterium (D2) lamp (Standard), tungsten (W) lamp (option)

    Flow Cell

    8 µL (10mm, TC), 12 MPa

    Option cell

    Conventional: 10 µL (10 mm, TC),
    Semi-micro: 2.5 µL (5 mm, TC)

    Read more
    Show Less

    Miscellaneous

    Dimensions

    W410×H605×D500 mm

    Weight

    67 kg

    Available pH range

    1 to 13

    Materials for parts in contact with liquids

    Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby,
    sapphire, Hastelloy® C, GFP, ceramic, PFA, quartz, PPS

    Workstation

    LabSolutions LC/GC Ver. 5.92 or later,
    LabSolutions DB/CS Ver. 6.82 SP1 or later (Incompatible with LCsolution )

    Read more
    Show Less