Language
Español
English
0

    Nexera series - Configuration

    Ultra High Performance Liquid Chromatograph

    System configuration

    System configuration

    Detector
    Pump
    Autosampler
    Mobile Phase Monitor
    System Controller
    Degassing Unit
    Column Oven
    Plate Changer

    System Controller

    SCL-40

     
     
     

    CBM-40

      SCL-40 CBM-40 CBM-40lite
    Monitor Touch panel
    LabSolutions™
    Web monitor
    LabSolutions
    Web monitor
    LabSolutions
    Web monitor
    Connectable unit
     
    Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Number of connectable units 8 (Using option: 12) 4 (Excluding built-in solvent delivery unit)
    Event input/output Input: 1, output: 2
    Analog board
     
    Up to two channels (option) Up to one channel (option) ̶
    Communication Ethernet
    Reservoir tray Built-in ̶
    Dimensions, weight W 260×D 500×H 140 mm、6 kg W 260×D 500×H 72 mm、5 kg ̶
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 50 VA, 50/60 Hz Supplied from solvent delivery unit

    Solvent Delivery Pump

     

    LC-40B XR

     

    LC-40B X3

      LC-40D
     
    LC-40D XR
    LC-40B XR
    LC-40D XS
    LC-40D XSi
    LC-40D X3
    LC-40B X3
    Pumping method Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure 44 MPa 70 MPa 105 MPa 130 MPa
    Flow rate settings range 0.0001-5.0000 mL/min
    (1.0-44 MPa)
    5.0001-10.0000 mL/min
    (1.0-22 MPa)
    0.0001-3.0000 mL/min
    (1.0-70 MPa)
    3.0001-5.0000 mL/min
    (1.0-44 MPa)
    5.0001-10.0000 mL/min
    (1.0-22 MPa)
    0.0001-3.0000 mL/min
    (1.0-105 MPa)
    3.0001-5.0000 mL/min
    (1.0-80 MPa)
    5.0001-10.0000 mL/min
    (1.0-22 MPa)
    0.0001-3.0000 mL/min
    (1.0-130 MPa)
    3.0001-5.0000 mL/min
    (1.0-80 MPa)
    5.0001-10.0000 mL/min
    (1.0-22 MPa)
    Flow rate accuracy ≤± 1% or ± 2 μL/min, whichever greater
    (under specified conditions)
    ≤± 1% (under specified conditions)
    Flow rate accuracy ≤ 0.06% RSD or 0.02 minSD, whichever greater
    Gradient mode
     
    High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR) High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient High-pressure gradient (2 solvents (LC-40B X3 standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D X3)
    Gradient range of set concentrations 0 to 100% (0.1% step)
    Gradient range of set concentrations ± 0.5% (under specified conditions)
    Wetted materials SUS316L, Hastelloy® C, PEEK, PTFE, Sapphire, Ruby SUS316L, Hastelloy C, PEEK, PE, Sapphire, Ruby
    Titanium、Titanium alloys、Nickel alloys、PEEK、PTFE、Ruby、Sapphire、Hastelloy C、PE (LC-40D XSi)
    Available pH range 1 to 14
    Automatic rinsing kit Option Standard equipment
    Degassing unit 1 unit connectable LC-40D XR:1 unit connectable LC-40B XR: 2 units connectable 1 unit connectable
    LC-40D X3:
    1 unit connectable
    LC-40B X3:
    pre-installed (5 port builtin), 1 unit connectable
    Dimensions  W 260×D 500×H 140 mm LC-40D X3:
    W 260×D 500×H 140 mm
    LC-40B X3:
    W 260×D 500×H 210 mm
    Weight 10 kg LC-40D XR:10 kg
    LC-40B XR:13 kg
    12 kg LC-40D X3:12 kg
    LC-40B X3: 21 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100-240 V、50/60 Hz
    150 VA LC-40D XR:150VA
    LC-40B XR:180 VA
    150 VA LC-40D X3:150 VA
    LC-40B X3:180 VA

    Degassing Unit

    DGU-403

     

      DGU-403 DGU-405
    Number of degassed solvents 3 5
    Degassed flow line capacity 400 μL/1 line
    Dimensions , weight W 260×D 500×H 72 mm、4 kg
    Operating temperature range 4 to 35°C
    Power supply Supplied from solvent delivery unit

    Autosampler

    SIL-40C XR

     

      SIL-40
    SIL-40C
    SIL-40 XR
    SIL-40C XR
    SIL-40C XS
    SIL-40C XSi
    SIL-40C X3
    Injection method
     
    Total-volume Injection (standard), loop injection (optional, except SIL-40C XSi)
    Allowable
    maximum pressure
    44 MPa 80 MPa 105 MPa 130 MPa
    Injection volume 0.01-100 μL 0.01-50 μL
    0.01 to 2000 μL (optional, except SIL-40C XSi)
    Injection volume
    accuracy
    ≤ ± 1% (5 μL injection, n = 20)
    Linearity ≥ 0.9999
    Injection cycle time ≤ 6.7 seconds (under specified conditions)
    Samples for processing   288( microtiter plate, 96 well × 3 plates)
      1152( microtiter plate, 384 well × 3 plates)
    252( 1 mL sample vial, 84 × 3 plates)
      162( 1.5 mL sample vial, 54 × 3 plates)
    84( 4 mL sample vial, 28 × 3 plates)
      36( 10 mL sample vial, 12 × 3 plates)
         72( 1.5 mL micro tube, 24 × 3 plates)) 
    Injection volume reproducibility
     
    RSD≦1.0% (0.5-0.9 μL)
    RSD≦0.5% (1.0-1.9 μL)
     RSD≦0.25%( 2.0-4.9 μL)
               RSD≦0.15%( More than 5.0 μL)
           RSD<0.5% (typically, 0.5 μL, except SIL-40C XSi)
             RSD<0.25%( typically, 1.0 μL, except SIL-40C XSi)
    Carryover
     
    ≤ 0.0025% (without rinse)
    ≤0.0005%
    (with rinse, typically)
    (under specified conditions)
    ≤ 0.0015%(without rinse)
    ≤ 0.0003%(with rinse, typically))
    (under specified conditions)
    Dip rinsing outside the needle
    and injection port rinsing
     
    Standard equipment
    Pumping rinse
    outside the needle
    Option Standard equipment
    Internal rinsing (3 dil) Option Option(SIL-40C XS)
    Standard equipment (SIL-40C XSi)
    Standard equipment
    Sample cooler
     
    SIL-40:None
    SIL-40C:Standard equipment
    (Air-circulation temperature
    control type)
     
    SIL-40 XR:None
    SIL-40C XR:Standard equipment
    (Air-circulation temperature
    control type)
     
    Standard equipment
    (Air-circulation temperature control type)
    Sample cooler
    temperature setting range
     
    4 to 45°C
    (Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)
    Sample cooler
    temperature accuracy
     
    ±2 ℃(sensor position ± 0.5°C)
    Wetted material SUS316L, DLC, PEEK, GFP, PTFE, FEP, ETFE, sapphire, ceramics, PPS, FFKM
    PEEK、ceramics、PTFE、ETFE、FEP、GFP、sapphire、PPS、FFKM (SIL-40C XSi)
    Available pH range 1 to 14
    Dimensions  W 260×D 500×H 280 mm(SIL-40C/40C XR/40C XS/40C XSi/40C X3Protrusion adds 140 mm to the depth)
    Weight SIL-40:17 kg
    SIL-40C:24 kg
     
    SIL-40 XR:17 kg
    SIL-40C XR:24 kg
    24 kg
    Operating
    temperature range
    4 to 35 ℃
    Power supply Cooler model AC 100-240 V, 50/60 Hz
    Non cooler model SIL-40:150 VA
    SIL-40C:400 VA
    SIL-40 XR:150 VA
    SIL-40C XR:400 VA
     

    Column Oven

    CTO-40C

    CTO-40S

      CTO-40C CTO-40S
    Temperature control type Forced air circulation
    Cooling Method Electronic cooling
    Temperature control range Room temperature − 10°C to 100°C Room temperature − 10°C to 85°C
    Temperature accuracy ±0.5 ℃ ±0.8 ℃
    Temperature precision 0.1 ℃(±0.05 ℃) 0.2 ℃(±0.1 ℃)
    Containable column size and
    number
    Up to 250 mm L. column × 6
    or 300 mm L. column × 3
    Up to 100 mm L. column × 6
    or 300 mm L. column × 3
    Dimensions, weight W 260×D 500×H 415 mm、21 kg W 130×D 500×H 553 mm、15 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100~120 V/AC 220-240 V(Automatic switching)
    400 VA、50/60 Hz
    AC 100-240 V、300 VA、50/60 Hz

    Detector

    SPD-40V

      SPD-40 SPD-40V
    Light source  Deuterium (D2) lamp Deuterium (D2) lamp, tungsten lamp
    Wavelength range 190 to 700 nm 190 to 1000 nm
    Bandwidth 8 nm
    Wavelength accuracy ≤ ± 1 nm±1 nm
    Wavelength reproducibility ≤ ± 0.1 nm
    Drift ≤ 0.1 × 10−3 of AU/h (under specified conditions)
    Noise 1 Wavelength mode: ≤ 4.0 × 10−6 AU, 2 Wavelength mode: ≤ 10.0 × 10−6 AU (under specified conditions)
    Linearity 2.5 AU (under specified conditions)
    Standard flow cell Optical path length: 10 mm, Cell volume: 12 μL, Pressure: 12 MPa
    Material of wetted parts: SUS316L, PFA, quartz, PEEK
    Cell temperature control range 19 to 50°C, 1°C Step
    Optional flow cell
     
    UHPLC cell (optical path length: 10 mm, cell volume: 8 μL, equipped with temperature control function)
    Semi-micro cell (optical path length: 5 mm, cell volume: 2.5 μL, equipped with temperature control function)
    Conventional cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function)
    Inert cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function)
    UHPLC inert cell (optical path length:10 mm、cell volume:8 μL、equipped with temperature control function)
    Low-diffusion inert cell (optical path length:5 mm、cell volume:2.5 μL、quipped with temperature control function)
    Preparative cell (optical path length: 0.1/0.2/0.5 mm, cell volume: 0.8/1.6/4.0 μL)
    Micro flow cell (optical path length: 3 mm, cell volume: 0.21 μL)
    Maximum pressure cell (optical path length: 10 mm, cell volume: 12 μL)
    Available pH range 1 to 13 (Cell quartz might be damaged by a mobile phase of pH >10.)
    Dimensions , weight W 260 × D 500 × H 140 mm, 11 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 150 VA, 50/60 Hz

    Photodiode Array Detector

    SPD-M40

      SPD-M40
    Light source Deuterium (D2) lamp, Tungsten lamp
    Number of diode elements 1024
    Wavelength range 190 to 800 nm
    Wavelength accuracy ≤ ±1 nm
    Wavelength reproducibility ≤ ±0.1 nm
    Slit width 1.2 nm、8 nm
    Spectral resolution ≤ ± 1.4 nm
    Drift ≤ 0.4×10-3 AU/h (under specified conditions)
    Noise ≤ 4.5 × 10−6 AU (under specified conditions)
    Linearity 2.5 AU (under specified conditions)
    Standard flow cell Optical path length: 10 mm, Cell volume: 12 μL, Pressure: 12 MPa
    Material of wetted parts: SUS316L, PFA, quartz, PEEK
    Cell temperature control range 19 to 50°C, 1°C Step
    Optional flow cell
     
    UHPLC cell (optical path length: 10 mm, cell volume: 8 μL, equipped with temperature control function)
    Semi-micro cell (optical path length: 5 mm, cell volume: 2.5 μL, equipped with temperature control function)
    Conventional cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function)
    Inert cell (optical path length: 10 mm, cell volume: 12 μL, equipped with temperature control function)
    UHPLC inert cell (optical path length:10 mm、cell volume:8 μL、equipped with temperature control function)
    Low-diffusion inert cell (optical path length:5 mm、cell volume:2.5 μL、quipped with temperature control function)
    Preparative cell (optical path length: 0.1/0.2/0.5 mm, cell volume: 0.8/1.6/4.0 μL, equipped)
    Micro flow cell (optical path length: 3 mm, cell volume: 0.21 μL)
    Maximum pressure cell (optical path length: 10 mm, cell volume: 12 μL)
    Available pH range 1 to 13 (Cell quartz might be damaged by a mobile phase pH >10.)
    Dimensions , weight W 260 × D 500 × H 140 mm, 10 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 180 VA, 50/60 Hz

    Capillary cell type Photodiode Array Detector

      SPD-M30A
    Light source Deuterium (D2) lamp
    Number of diode elements 1024
    Wavelength range 190 to 700 nm
    Wavelength accuracy ≤ ±1 nm
    Wavelength reproducibility ≤ ±0.1 nm
    Slit width 1 nm, 8 nm
    Spectral resolution ≤ 1.4 nm
    Drift ≤ 4.0 × 10−6 AU (under specified conditions)
    Noise ≤ 4.0 × 10−6 AU (under specified conditions)
    Linearity 2.0 AU (under specified conditions)
    Cell Standard cell: Optical path length: 10 mm, Capacity: 1 μL, Pressure: 8 MPa
    Optional high-sensitivity cell: Optical path length: 85 mm, Capacity: 9 μL, Pressure: 8 MPa
    Optional low-dispersion cell PLUS: Optical path length: 10 mm, Capacity: 1 µL, Pressure: 8 MPa
    Dimensions, weight W 260×D 500×H 140 mm,12 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 150 VA, 50/60 Hz

    Spectrofluorometric Detector

      RF-20A RF-20AXS
    Light source
     
    Xenon lamp Xenon lamp
    Low-pressure mercury lamp
    (to check wavelength accuracy)
    Wavelength range 200 to 650 nm 200 to 750 nm
    Spectral bandwidth 20 nm
    Wavelength accuracy
     
    ±2 nm
     
    Wavelength precision
    ±0.2 nm
    S/N
     
    Water Raman peak S/N ≥ 200
    Low background S/N ≥ 9000
    Water Raman peak S/N ≥ 2000
    Low background S/N ≥ 12000
    Range of cell temperature control ̶ Room temperature − 10°C to 40°C, 1°C step
    S/N
    Water Raman peak S/N ≥1200
    Low background S/N ≥ 9000
    Water Raman peak S/N ≥ 2000
    Low background S/N ≥ 12000
    Cell Standard conventional cell: volume 12 μL, maximum pressure 2 MPa
    Optional semi-micro cell: volume 3 μL, maximum pressure 2 MPa
    Option inert cell: volume 12 µL, maximum pressure 2 MPa
    Function Simultaneous measurement of four wavelengths, Wavelength scanning
    Dimensions  W 260×D 500×H 210 mm
    Weight 17 kg 18 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 400 VA, 50/60 Hz

    Differential Refractive Index Detector

      RID-20A
    Measurement range 1 to 1.75 RIU
    Noise ≤ 2.5×10-9 RIU
    Drift ≤ 1×10-7 RIU/h
    Range A mode: 0.01 × 10−6 to 500 × 10−6 RIU
    P, L-mode: 1 × 10−6 to 5000 × 10−6 RIU
    Response
    0.05 to 10 sec, 10 steps
    Polarity – Change Available
    Zero adjustment Auto zero, Optical zero, Fine zero
    Maximum flow rate
    20 mL/min (150 mL/min in option)
    Range of cell temperature control 30 to 60°C
    Cell Volume 9 μL, Maximum pressure 2 MPa (equipment : 0.4 MPa)
    Dimensions, weight W 260 × D 420 × H 140 mm, 12 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100–240 V, 150 VA, 50/60 Hz

    Conductivity Detector

      CDD-10Avp
    Cell volume 0.25 μL
    Cell constant 25 μS·cm⁻¹
    Material of wetted parts PEEK, SUS316
    Maximum use pressure 2.9 MPa(30 kgf/cm2)
    Response
    0.05 to 10 s, 10 steps
    Zero adjustment Auto-zero function, Baseline-shifting function
    Dimensions, weight W 260 × D 420 × H 140 mm, 6 kg
    Operating temperature range
    4 to 35°C
    Power supply AC 100–240 V, 250 VA, 50/60 Hz

    Evaporative Light-Scattering Detector

      ELSD-LT II ELSD-LT III
    Nebulizing method Siphon Splitting Siphon Splitting
    Light source LED Semiconductor laser
    Detector Photomultiplier Photodiode
    Temperature Setting Range Room temperature + 5°C to 80°C Room temperature to 100 °C
    Nebulizer Gas
    Nitrogen or air* Nitrogen or air*
    Gas flow rate, gas pressure Up to 3.0 mL/min, up to 450 kPa Up to 3.0 mL/min, up to 450 kPa
    Mobile Phase Flow Rate
    (Standard Nebulizer)
    0.2 to 2.5 mL/min 0.2 to 2 mL/min
    Analog output
    0 to 1 V 0 to 1 V
    Dimensions W 260×D 550×H 450 mm W 250×D 530×H 330 mm
    weight 20 kg 15.5 kg
    Operating temperature range 5 to 40°C 4 to 35°C
    Operation humidity range ≤ 80% (Room temperature + 5°C to 31°C),
     ≤ 50% (Room temperature + 31°C to 40 °C)
    ≦80% (4~31℃),
    ≦50% (31~35℃)
    Power supply AC 100 V, 210 VA, 50/60 Hz 100 to 240 VAC, 120 VA, 50/60 Hz

    * Supply gas at a pressure of about 350 kPa.
       An air compressor may also be used.
       A filter (P/N: 228-45528-92) is also available for filtering out moisture and other matter from the compressor.
       For more details, contact a Shimadzu sales representative. 

    Mobile Phase Monitor  MPM-40 ANALYTICAL INTELLIGENCE

    Mobile Phase Monitor
    • Monitor mobile phase levels remotely from a PC or smartphone
    • Avoid running out of mobile phase or rinse solution mid-analysis
    • Fits a range of bottle sizes from 1 to 5L

     

        Description
    Mobile Phase Monitor (Controller)   This is a controller to connect the bottle holders.
    It can connect two reserve trays of bottle holders.
    Mobile Phase Monitor (1L bottle holder)

    1L bottle holder

    One holder can accommodate two 1L bottles.
    Mobile Phase Monitor (Large bottle holder)

    Large bottle holder

    One holder can accommodate one large bottle.

    ※ The large bottle holder can hold bottles up to a diameter of approx. 180 mm.
    ※ To change between the two holder types, a field engineer will need to carry out the modifications.

     

    Versatile layout can accommodate different bottle combinations

    Two trays of bottles can be monitored from one MPM controller.

     
     
     
    ANALYTICAL INTELLIGENCE

    ANALYTICAL INTELLIGENCE

    Automated support functions utilizing digital technology, such as M2M, IoT, and Artificial Intelligence (AI), that enable higher productivity and maximum reliability.

     

     

     

    pH monitor

    pHM-40

    pHM-40

      pHM-40
    Available pH range 1 to 14
    Precision pH ±0.1 (under specified conditions)
    Drift pH ±0.1 /10 h (under specified conditions)
    Maximum flow rate 10 mL/min
    Allowable maximum pressure 0.1 MPa
    Cell volume approx. 80 μL
    Available liquid temperature 4 to 60 °C
    Wetted materials glass, PEEK, PCTFE, silicon
    Calibration point 1 to 5 point is settable.
    Dimensions [mm], weight W 130 × D 393 × H 206, 6 kg
    Power supply DC 5 V, 5 VA

     

    Plate Changer

      PLATE CHANGER
    Samples for processing
    (includes two plates of autosampler)
    1 PLATE CHANGER
    1536 (microtiter plate, 96 well × 16 plates)
    864 (deep-well plate, 96 well × 9 plates)
    6144 (microtiter plate, 384 well × 16 plates) 
    3456 (deep-well plate, 384 well × 9 plates)
    756 (1 mL sample vial, 84 × 9 plates)
    486 (1.5 mL sample vial, 54 × 9 plates)
    252 (4 mL sample vial, 28 × 9 plates)
    108 (10 mL sample vial, 12 × 9 plates)
    3 PLATE CHANGERS
    4224 (microtiter plate, 96 well × 44 plates)
    2208 (deep-well plate, 96 well × 23 plates)
    16896 (microtiter plate, 384 well × 44 plates)
    8832 (deep-well plate, 384 well × 23 plates)
    1932 (1 mL sample vial, 84 × 23 plates)
    1242 (1.5 mL sample vial, 54 × 23 plates)
    644 (4 mL sample vial, 28 × 23 plates)
    276 (10 mL sample vial, 12 × 23 plates)
    Sample cooler
    temperature setting range
    Air-circulation temperature control type, 4 to 45°C
    (Room temperature needs to be less than 30°C and humidity needs to be less than 70% to set 4°C)
    Dimensions, weight W 170 × D 500 × H 560 mm (Protrusion adds 140 mm to the depth), 26 kg
    Operating temperature range 4 to 35°C
    Power supply AC 100−240 V, 400 VA, 50/60 Hz
     
     

    High-Pressure Gradient Using Dual Pump, Photodiode Array Detector

    P/N Description Q'ty
    228-65500-46/58 SCL-40 1
    228-65078-46/58 LC-40B X3 1
    228-65120-46/58 SIL-40C X3 1
    228-65201-46/58 CTO-40S 1
    228-65302-46/58 SPD-M40 1
    228-64724-41 UHPLC Cell 1
    228-72652-42 Mixer MR 40 μL
    (with a mixer recognition device)
    1
    228-70254-42 Piping Kit B, for High-Pressure GE ID0.1 1
    228-70247-43 Nexera Series Cable Installation Kit, C 1
    Power Cord, Length: 2.5 m 5
    228-57647-43 Hand Tool Kit 1

    High-Pressure Gradient Using Dual Pump, Plate Changer, Photodiode Array Detector

    P/N Description Q'ty
    228-65500-46/58 SCL-40 1
    228-65078-46/58 LC-40B X3 1
    228-65120-46/58 SIL-40C X3 1
    228-65110-46/58 Plate Changer 1
    228-65201-46/58 CTO-40S 1
    228-65302-46/58 SPD-M40 1
    228-64724-41 UHPLC Cell 1
    228-72652-42 Mixer MR 40 μL
    (with a mixer recognition device)
    1
    228-70254-42 Piping Kit B, for High-Pressure GE ID0.1 1
    228-70247-43 Nexera Series Cable Installation Kit, C 1
    Power Cord, Length: 2.5 m 6
    228-57647-43 Hand Tool Kit 1

    High-Pressure Gradient Using Two Pumps, Photodiode Array Detector

    P/N Description Q'ty
    228-65500-46/58 SCL-40 1
    228-65076-46/58 LC-40D X3 2
    228-65019-58 DGU-405 1
    228-65120-46/58 SIL-40C X3 1
    228-65202-46/58 CTO-40C 1
    228-65302-46/58 SPD-M40 1
    228-64724-41 UHPLC Cell 1
    228-72652-42 Mixer MR 40 μL
    (with a mixer recognition device)
    1
    228-70254-42 Piping Kit B, for High-Pressure GE ID0.1 1
    228-70247-43 Nexera Series Cable Installation Kit, C 1
    Power Cord, Length: 2.5 m 6
    228-57647-43 Hand Tool Kit 1

    High-Pressure Gradient Using Dual Pump, UV-VIS Detector

    P/N Description Q'ty
    228-65500-46/58 SCL-40 1
    228-65078-46/58 LC-40B X3 1
    228-65120-46/58 SIL-40C X3 1
    228-65202-46/58 CTO-40C 1
    228-65313-46/58 SPD-40V 1
    228-64724-42 UHPLC Cell 1
    228-72652-42 Mixer MR 40 μL
    (with a mixer recognition device)
    1
    228-70254-42 Piping Kit B, for High-Pressure GE ID0.1 1
    228-70247-44 Nexera Series Cable Installation Kit, D 1
    Power Cord, Length: 2.5 m 5
    228-57647-43 Hand Tool Kit 1

    Low-Pressure Gradient, Photodiode Array Detector

    P/N Description Q'ty
    228-65500-46/58 SCL-40 1
    228-65076-46/58 LC-40D X3 1
    228-65016-46/58 Low-Pressure Gradient Kit 1
    228-65019-58 DGU-405 1
    228-65120-46/58 SIL-40C X3 1
    228-65201-46/58 CTO-40S 1
    228-65302-46/58 SPD-M40 1
    228-64724-41 UHPLC Cell 1
    228-65020-41 Mixer Installation Kit, MR 40 μL LPGE (with a mixer recognition device) 1
    228-70247-42 Nexera Series Cable Installation Kit, B 1
    Power Cord, Length: 2.5 m 5
    228-57647-43 Hand Tool Kit 1
    228-36573 1/16"NYLON COUPLING 2
    228-56687-42 SF NUT,1/8 4
    228-56687-43 SF FERRULE,1/8 4

    High Performance Liquid Chromatography Nexera™ lite, ELSD-LT III

    P/N Description Q'ty
    LC-40D 2
    PEEK Tube, 0.13mmIDx3000mm 1
    Shodex HILICpak VG-50 4E(250 × 4.6 mm I.D., 5 μm) 1
    SHIMADZU LabTotal™ for LC 1.5 mL, Glass A 1
    LC LabSolutions 1
    LabSolutions LC Workstation Single 1
    TOOL KIT 1 1
    CORD,UC-975-N01 5
    Optical Fiber Cable Assy, Plastic, for Nexera Series 1
    REGULATOR,W/FILTER 1
    PEEK Male Nut 2
    CBM-40 lite 1
    PIPING KIT A, for HPGE ID0.3 1
    MR 180μL II 1
    RESERVOIR TRAY 1
    ELSD-LT III 1
    CTO-40C 1
    VIAL PLAT 1.5mL 2
    SIL-40C 1
    DGU-403 1

    Nexera™ XR High Performance Liquid Chromatograph

    P/N Description Q'ty
    LC-40B XR 1
    Mobile Phase Monitor (Contoller) 1
    LC LabSolutions 1
    LabSolutions LCGC,DB Multi LC-PDA 1
    SYSTEM BASE KIT B 2
    LOCKING PLATE 1
    BELT 1
    MPM Bottle Holder Kit,Eeathquake-R 1
    CMD CABLE 1
    Mobile Phase Monitor AC Adapter 1
    Mobile Phase Monitor(bottle holder) 2
    Glass Reservoir Bottle, w/Cap, 5pcs 1
    SCL-40 1
    TOOL KIT 1
    POWER OUTLET UNIT 1
    Optical Fiber Cable Assy, Plastic, for Nexera Series, C 1 CORD,UC-975-N01 5
    PIPING KIT B, for HPGE ID0.1 1
    MR 180μL II 1
    SPD-M40 1
    CTO-40C 1
    VIAL PLAT 1.5mL 2
    SIL-40CXR 1
    DGU-403 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    SCL-40 1
    LC-40B XR 1
    DGU-403 1
    SIL-40C XR
    CTO-40C 1
    SPD-M40 1
    CRB-40 1
    D582/F LCsolution Ver.1.25 AD 1
    GLP 0.1φX2M 1
    GLP 0.5φX1M 1
    GLP 0.5φX5M 1
    FEMALE T 1.6FT316L-0.4 1
    PEEK Male Nut, 5pcs/pack 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    LC-40D 2
    SCL-40 1
    DGU-403 1
    SIL-40C 1
    CTO-40C 1
    ELSD-LTIII 1
    Imtakt Presto FF-C18(150 mm×4.6 mmI.D.;2 μm) 1
    Labtotal Vial KIT 1
    MR20 μL 1
    Piping kit 0.3 mm 1
    Cable Assy 1
    LC LabSolutions Single LC 1
    REGULATOR,W/FILTER 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    CBM-40 1
    Bottle, 250ml 2
    Shim-pack GIST-HP C18-AQ 1
    SCR-102(H) 1
    Shim-pack SCR-102H 2
    TOOL KIT 1
    FITTING SET 20A 2
    CABLE,HFBR3600-1-021 4
    COLUMN CALMP B8 1
    Vial plate 1.5mL 2
    WASHING PUMP SEAL KIT 2
    LC-40D 2
    RESERVOIR TRAY 1
    LabsolutionsLCGC,Single 1
    SPD-40 1
    TUBING ASSY 1
    CDD-10Avp 1
    CTO-40C 1
    SIL-40C 1
    DGU-403 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    LC-40D 2
    MPM Bottle Holder Kit,Eeathquake-R 1
    Shim-pack Scepter C18-120 A 1
    Shim-pack VP-Phenyl 1
    SHIMADZU Vials, TORAST-H Glass Vial, 1.5mL 1
    LC LabSolutions 1
    LabSolutions LCGC,DB Single LC 1
    SYSTEM BASE KIT B 1
    LOCKING PLATE 1
    BELT 1
    CMD CABLE 1
    CBM-40Lite 1
    Mobile Phase Monitor AC Adapter CN 1
    Mobile Phase Monitor(bottle holder) 1
    Mobile Phase Monitor (Contoller) 1
    Glass Reservoir Bottle, w/Cap, 5pcs 1
    RESERVOIR TRAY 1
    SPD-40V 1
    CTO-40C 1
    SIL-40C 1
    DGU-403 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    CBM-40 1
    Shim-pack SCR-102(H) 2
    LC PC LabSolutions PC 1
    TOOL KIT 1
    CORD,UC-975-N01 1
    PEEK Male Nut 1
    PEEK Tube, 0.25mmIDx3000mm 1
    PEEK Tube, 0.13mmIDx3000mm 1
    FITTING SET 20A 1
    LabSolutions LCGC,DB Single LC 1
    Reagent Kit for organic acid analysis system 1
    GUARD COLUMN FOR SCR-102H 1
    TUBING ASSY 1
    LC-40D 2
    CABLE,HFBR3600-1-021 1
    RESERVOIR TRAY 1
    CDD-10Avp 1
    COLUMN CALMP B8 2
    CTO-40C 1
    Vial plate 1.5mL 1
    SIL-40C 1
    DGU-403 1
    Bottle, 250ml 1
    WASHING PUMP SEAL KIT 2

    High Performance Liquid Chromatography

    P/N Description Q'ty
    SCL-40 1
    LC-40B XR 1
    DGU-405 1
    SIL-40CXR 1
    CTO-40S 1
    SPD-M40 1
    MR 180μLⅡ 1
    LC WORKSTATION 1
    LC WORKSTATION Labsolution 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    SCL-40 1
    LC-40B XR 1
    DGU-403 1
    SIL-40CXR 1
    Vial plate 1.5mL 1
    CTO-40S CN 1
    SPD-M40 1
    MR 180 μL II 1
    PIPING KIT B, for HPGE ID0.1 1
    Optical Fiber Cable Assy, Plastic, for Nexera Series,C 1
    UC-975-N01 2.5m 1
    LabSolutions LCGC,DB Multi LC-PDA 1

    High Performance Liquid Chromatography

    P/N Description Q'ty
    SCL-40 1
    LC-40D XR 1
    DGU-40A5 1
    LPGE UNIT 1
    SIL-40 1
    CTO-40C 1
    SPD-M40A 1
    LC WORKSTATION 1
    LC WORKSTATION LabSolutions 1
    SEMI MICRO TEMP CONT CELL 1
    Mixer (300 uL) 1

    Do you live in the United States?