0

    Nexera series - Specs

    Ultra High Performance Liquid Chromatograph

    Nexera X3

    Pressure

    130MPa

    Approx. particle size (*)

    sub2 μm-3μm

    I.D. (*)

    1-3mm

    Dual pump

    Multi-rinse

    Installed as standard

    Key features

    Shimadzu’s flagship UHPLC model. This high-pressure gradient model minimizes gradient delay volume and is ideal for very high sample numbers or as the front end of an LC/MS system. It is also compatible with low-pressure gradients and mobile phase blending.

    *To achieve maximum functionality of both instruments and columns, we recommend using columns with the particle size and dimensions suggested. However, as long as the maximum pressure of the instrument and columns are not exceeded, other columns can be used.
    Piping adjustments may be necessary depending on the analysis conditions. Please contact a Shimadzu sales representative for details.

    Read more
    Show Less

    Shimadzu’s flagship UHPLC model. This high-pressure gradient model minimizes gradient delay volume and is ideal for very high sample numbers or as the front end of an LC/MS system. It is also compatible with low-pressure gradients and mobile phase blending.

     Nexera X3 High Pressure Gradient 
      Model Name Features
    System controller SCL-40 Color screen with touch panel and system controller with reservoir tray
    Number of connectable units : 8 (Using option: 12)
    Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Solvent Delivery Pump LC-40B X3
    Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure : 130 MPa
    Flow rate settings range : 0.0001 – 3.0000 mL/min (1.0 – 130 MPa), 3.0001 – 5.0000 mL/min (1.0 – 80 MPa), 5.0001 – 10.0000 mL/min (1.0 – 22 MPa)
    Flow rate accuracy :  ≦± 1% (under specified conditions)
    Gradient mode : High-pressure gradient (2 solvents (LC-40B X3 standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D X3)
    Gradient Mixer
    Autosampler SIL-40C X3
    Injection method : Total-volume Injection (standard), loop injection (optional)
    Allowable maximum pressure : 130 MPa
    Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional)
    Injection volume accuracy :  ≦± 1% (5 μL injection, n = 20)
    Injection cycle time :  ≦6.7 seconds (under specified conditions)
    Column Oven CTO-40S Temperature control type : Forced air circulation
    Cooling Method : Electronic cooling
    Temperature control range : Room temperature − 10°C to 85°C
    Temperature accuracy : ± 0.8°C
    Temperature precision : ± 0.1°C
    UV-VIS Detector SPD-M30A
    Light source : Deuterium (D2) lamp
    Number of diode elements : 1024
    Wavelength range : 190 to 700 nm
    Wavelength accuracy :  ≦± 1 nm
    Wavelength reproducibility :  ≦± 0.1 nm
    Mobile phase monitor
    Software Labsolutions LC/GC Multi LC-PDA


     

    Nexera XS

    Pressure

    105MPa

    Approx. particle size (*)

    sub 2μm-3μm

    I.D. (*)

    1-3mm

    Dual pump

    -

    Multi-rinse

    Optional

    Key features

    A highly flexible UHPLC model compatible with both high and low gradients. Not only does this model significantly reduce analysis time, but the use of plate-changers enables efficient analysis of a large number of samples.

    *To achieve maximum functionality of both instruments and columns, we recommend using columns with the particle size and dimensions suggested. However, as long as the maximum pressure of the instrument and columns are not exceeded, other columns can be used.
    Piping adjustments may be necessary depending on the analysis conditions. Please contact a Shimadzu sales representative for details.

    Read more
    Show Less

    A highly flexible UHPLC model compatible with both high and low gradients. Not only does this model significantly reduce analysis time, but the use of plate-changers enables efficient analysis of a large number of samples.

     Nexera XS Low Pressure Gradient 
      Model Name Features
    System controller SCL-40 Color screen with touch panel and system controller with reservoir tray
    Number of connectable units : 8 (Using option: 12)
    Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Solvent Delivery Pump LC-40D XS
    Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure : 105 MPa
    Flow rate settings range : 0.0001 – 3.0000 mL/min (1.0 – 105 MPa) 3.0001 – 5.0000 mL/min (1.0 – 80 MPa) 5.0001 – 10.0000 mL/min
    (1.0 – 22 MPa)
    Flow rate accuracy :≦± 1% (under specified conditions)
    Gradient mode : High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient
    Gradient Mixer
    Degassing Unit DGU-405 Number of degassed solvents : 5
    Degassed flow line capacity : 400 μL/1 line
    Operating temperature range : 4 - 35°C
    Autosampler SIL-40C XS
    Injection method : Total-volume Injection (standard), loop injection (optional)
    Allowable maximum pressure : 105 MPa
    Injection volume : 0.1 - 50 μL, 0.1 - 2000 μL (optional)
    Injection volume accuracy : ≦± 1% (5 μL injection, n = 20)
    Injection cycle time : ≦6.7 seconds (under specified conditions)
    Column Oven CTO-40S Temperature control type : Forced air circulation
    Cooling Method : Electronic cooling
    Temperature control range : Room temperature − 10°C - 85°C
    Temperature accuracy : ± 0.8°C
    Temperature precision : ± 0.1°C
    UV-VIS Detector SPD-M30A Light source : Deuterium (D2) lamp
    Number of diode elements : 1024
    Wavelength range : 190 to 700 nm
    Wavelength accuracy : ≦± 1 nm
    Wavelength reproducibility : ≦± 0.1 nm
    Mobile phase monitor
    Software Labsolutions LC/GC Multi LC-PDA


     

    Nexera XR

    Pressure

    70MPa

    Approx. particle size (*)

    2μm-3μm

    I.D. (*)

    2-4.6mm

    Dual pump

    Multi-rinse

    Optional

    Key features

    This UHPLC-like model is the new standard for the Shimadzu LC series. It is compatible with a wide range of analysis conditions from conventional HPLC analysis to ultra-high separation analysis using core-shell columns with a particle size of 3 μm and below.

    *To achieve maximum functionality of both instruments and columns, we recommend using columns with the particle size and dimensions suggested. However, as long as the maximum pressure of the instrument and columns are not exceeded, other columns can be used.
    Piping adjustments may be necessary depending on the analysis conditions. Please contact a Shimadzu sales representative for details.

    Read more
    Show Less

    This UHPLC-like model is the new standard for the Shimadzu LC series. It is compatible with a wide range of analysis conditions from conventional HPLC analysis to ultra-high separation analysis using core-shell columns with a particle size of 3 μm and below.

     Nexera XR High Pressure Gradient 
      Model Name Features
    System controller SCL-40 Color screen with touch panel and system controller with reservoir tray
    Number of connectable units : 8 (Using option: 12)
    Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Solvent Delivery Pump LC-40B XR
    Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure : 70 MPa
    Flow rate settings range : 0.0001 – 3.0000 mL/min (1.0 – 70 MPa)
    3.0001 – 5.0000 mL/min (1.0 – 44 MPa) 5.0001 – 10.0000 mL/min
    (1.0 – 22 MPa)
    Flow rate accuracy :  ≦± 1% or ± 2 μL/min, whichever greater
    (under specified conditions)
    Gradient mode : High-pressure gradient (2 solvents (LC-40B XR standard) or 3 solvents) Quaternary low-pressure gradient (Only available for LC-40D XR)
    Gradient Mixer
    Degassing Unit DGU-403 Number of degassed solvents : 3
    Degassed flow line capacity : 400 μL/1 line
    Operating temperature range : 4 to 35°C
    Autosampler SIL-40C XR
    Injection method : Total-volume Injection (standard), loop injection (optional)
    Allowable maximum pressure : 80 MPa
    Injection volume : 0.1 to 50 μL, 0.1 to 2000 μL (optional)
    Injection volume accuracy :  ≦± 1% (5 μL injection, n = 20)
    Injection cycle time :  ≦6.7 seconds (under specified conditions)
    Column Oven CTO-40S Temperature control type : Forced air circulation
    Cooling Method : Electronic cooling
    Temperature control range : Room temperature − 10°C to 85°C
    Temperature accuracy : ± 0.8°C
    Temperature precision : ± 0.1°C
    UV-VIS Detector SPD-M40 Light source :Deuterium (D2) lamp, Tungsten lamp
    Number of diode elements : 1024
    Wavelength range : 190 to 800 nm
    Wavelength accuracy :  ≦± 1 nm
    Wavelength reproducibility :  ≦± 0.1 nm
    Mobile phase monitor
    Software Labsolutions LC/GC Multi LC-PDA


     

    Nexera lite

    Pressure

    44MPa

    Approx. particle size (*)

    above 3μm

    I.D. (*)

    3mm-

    Dual pump

    -

    Multi-rinse

    Optional

    Key features

    A conventional HPLC model with excellent basic functions and many options for expansion. Cutting-edge AI functions increase the reliability of established analysis workflows. It can be adapted for all application systems.

    *To achieve maximum functionality of both instruments and columns, we recommend using columns with the particle size and dimensions suggested. However, as long as the maximum pressure of the instrument and columns are not exceeded, other columns can be used.
    Piping adjustments may be necessary depending on the analysis conditions. Please contact a Shimadzu sales representative for details.

    Read more
    Show Less

    A conventional HPLC model with excellent basic functions and many options for expansion. Cutting-edge AI functions increase the reliability of established analysis workflows. It can be adapted for all application systems.

     Nexera lite High Pressure Gradient 
      Model Name Features
    System controller SCL-40 Color screen with touch panel and system controller with reservoir tray
    Number of connectable units : 8 (Using option: 12)
    Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Solvent Delivery Pump LC-40D (2pcs)
    Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure : 44 MPa
    Flow rate settings range : 0.0001 – 5.0000 mL/min (1.0 – 44 MPa)
    5.0001 – 10.0000 mL/min (1.0 – 22 MPa)
    Flow rate accuracy :  ≦± 1% or ± 2 μL/min, whichever greater
    (under specified conditions)
    Gradient mode : High-pressure gradient (2 or 3 solvents) Quaternary low-pressure gradient
    Gradient Mixer
    Degassing Unit DGU-403 Number of degassed solvents : 3
    Degassed flow line capacity : 400 μL/1 line
    Operating temperature range : 4 to 35°C
    Autosampler SIL-40C
    Injection method : Total-volume Injection (standard), loop injection (optional)
    Allowable maximum pressure : 44 MPa
    Injection volume : 0.1 to 100 μL, 0.1 to 2000 μL (optional)
    Injection volume accuracy :  ≦± 1% (5 μL injection, n = 20)
    Injection cycle time :  ≦6.7 seconds (under specied conditions)
    Column Oven CTO-40C Temperature control type : Forced air circulation
    Cooling Method : Electronic cooling
    Temperature control range : Room temperature − 10°C to 100°C
    Temperature accuracy : ± 0.5°C
    Temperature precision : ± 0.05°C
    UV-VIS Detector SPD-M40 Light source : Deuterium (D2) lamp
    Number of diode elements : 1024
    Wavelength range : 190 to 700 nm
    Wavelength accuracy : ≦± 1 nm
    Wavelength reproducibility :  ≦± 0.1 nm
    Mobile phase monitor
    Software Labsolutions LC/GC Multi LC-PDA


     
     Nexera lite Low Pressure Gradient 
      Model Name Features
    System controller CBM-40 Monitor : LabSolutions Web monitor
    Connectable unit : Solvent delivery unit: max. 4, Autosampler: 1, Column oven: max. 4, Detector: max 2, etc.
    Number of connectable units : 8 (Using option: 12)
    Solvent Delivery Pump LC-40D
    Pumping method : Parallel-type double plunger (approx. 10 μL/1 stroke)
    Allowable maximum pressure : 44 MPa
    Flow rate settings range : 0.0001 – 5.0000 mL/min (1.0 – 44 MPa)
    5.0001 – 10.0000 mL/min (1.0 – 22 MPa)
    Flow rate accuracy :  ≦± 1% or ± 2 μL/min, whichever greater(under specified conditions)
    Gradient mode : High-pressure gradient (2 or 3 solvents Quaternary low-pressure gradient
    Gradient Mixer
    Degassing Unit DGU-405 Number of degassed solvents : 5
    Degassed flow line capacity : 400 μL/1 line
    Operating temperature range : 4 - 35°C
    Autosampler SIL-40C
    Injection method : Total-volume Injection (standard), loop injection (optional)
    Allowable maximum pressure : 44 MPa
    Injection volume : 0.1 to 100 μL, 0.1 - 2000 μL (optional)
    Injection volume accuracy : ≦± 1% (5 μL injection, n = 20)
    Injection cycle time : ≦6.7 seconds (under specified conditions)
    Column Oven CTO-40S Temperature control type : Forced air circulation
    Cooling Method : Electronic cooling
    Temperature control range : Room temperature ― 10°C - 85°C
    Temperature accuracy : ± 0.8°C
    Temperature precision : ± 0.1°C
    UV-VIS Detector SPD-M40 Light source :Deuterium (D2) lamp, Tungsten lamp
    Number of diode elements : 1024
    Wavelength range : 190 - 800 nm
    Wavelength accuracy :  ≦± 1 nm
    Wavelength reproducibility :  ≦± 0.1 nm
    Mobile phase monitor
    Software Labsolutions LC/GC Multi LC-PDA


     
    English
    English (United States)
    Español (Estados Unidos)