0

    IRTracer-100 - Specs

    Fourier Transform Infrared Spectrophotometer

    IRTracer-100

    Interferometer

    Michelson interferometer (30° incident angle)
    Equipped with Advanced Dynamic Alignment system
    Sealed interferometer with Automatic Dehumidifier

    Beam splitter

    Germanium-coated KBr for Middle IR (Standard)
    Germanium-coated CsI for Middle/Far IR (Optional)
    Silicon-coated CaF2 for Near IR (Optional)

    Light source

    High-energy ceramic for Middle/Far IR (Standard)
    Tungsten lamp for Near IR (Optional)

    Detector

    DLATGS detector with temperature control for Middle/Far IR (Standard)
    T2SL detector with liquid nitrogen cooling for Middle/Near IR (Optional)
    InGaAs detector for Near IR (Optional)

    Wavenumber range

    7,800 to 350 cm-1 (Standard)
    12,500 to 240 cm-1 (Optional)

    Resolution

    0.25, 0.5, 1, 2, 4, 8, 16 cm-1 (Middle/Far IR)
    2, 4, 8, 16 cm-1 (Near IR)

    Dimensions

    600 (W) x 665 (D) x 295 (H) mm

    Weight

    47 kg

    S/N ratio

    60,000:1 or higher

    Mirror speed

    Standard: 2, 2.8, 5, 9 mm/sec
    Optional: 10, 20, 30, 40 mm/sec (Rapid scan)