Standard Specifications
Turbo molecular pump | TMP-X3405LMTC | |
---|---|---|
Inlet flange | VG300 / ISO320B / VG350 | |
Outlet flange | KF40 | |
Cooling method | Water | |
Ultimate pressure (after baking) | 10-7 Pa order | |
Maximum allowable Argon gas flow rate (Note 1) |
800 mL/min (Note 2) (Pump's set temperature : 65 degrees C.) |
|
Maximum allowable inlet pressure (N2 continuous exhaust) (Note 1) |
40 Pa | |
Maximum allowable outlet pressure (N2 continuous exhaust) (Note 1) |
270 Pa | |
Pumping speed (Note 3) |
N2 | 3200 L/s |
Ar | 3100 L/s | |
H2 | 2150 L/s | |
Compression ratio | N2 | 2 x 108 |
H2 | 2 x 103 | |
Rated speed | 27600 rpm | |
Start-up time | 12 minutes or less | |
Mounting position | In any desired direction | |
Vibration level (by Shimadzu's method) | 0.05 µm or less (0-peak) | |
Recommended pumping speed of backing vacuum pump in case of gas purge | 2000 L/min or more | |
Display | LED | POWER / ROTATION / TEMP.CONTROL / ALARM |
Communication | Contact |
REMOTE (D-sub 25 pin female, Screw lock size UNC 4-40) Input : START / STOP / RESET / LOW SPEED Output : ROTATION / NORMAL / ACC. / BRAKE / REMOTE / TEMP.STABLE / ALARM / WARNING |
Serial | RS-232C / RS-485 | |
Speed variation | Speed is variable between 25 % and 100 % of the rated speed (set as 0.1 %). | |
Water | Flow rate | 3 L/min or more |
Pressure | 0.2 to 0.4 MPa | |
Temperature | 19 to 30 degrees C. (above dew point) | |
Admissible ambient magnetic field | Radial direction | 3 mT |
Axial direction | 15 mT | |
Input electric power | Voltage | Single phase 200 to 240 VAC ± 10 % (50 / 60 Hz ± 2 Hz) |
Maximum power | 2.1 kVA | |
Mass | 89 kg |
This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.