W-AD Series Analytical Balance

(Equipped Standard with Smart Automatic Doors and Touchless Sensors)

Model AP225W-AD AP135W-AD AP225WD-AD AP125WD-AD AP324W-AD AP224W-AD
Capacity 220g 135g 220g / 102g 120g / 52g 320g 220g
 Minimum Display  0.01mg 0.1mg / 0.01mg 0.1mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
95 to 220.00090g
(200g)
45 to 120.00090g
(100g)
95 to 320.0090g
(300g)
95 to 220.0090g 
(200g)
 Repeatability (at weighing capacity) 
(*2)
0.015mg (up  to 20g)
0.03mg (up to 100g)
0.05mg (up to 200g)
0.05mg 0.1mg / 0.05mg 0.1mg / 0.02mg 0.15mg 0.1mg
Repeatability
(for low loads) (*2, *3)
0.01mg
(for 5g load)
0.1mg
(for 20g load)
0.1mg
(for 10g load)
 Minimum Weight (*2, *3) 20mg (*2) 200mg
 Linearity (*2) ±0.1mg ±0.2mg / ±0.1mg ±0.2mg / ±0.05mg ±0.3mg ±0.2mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
 USB Host (Type A)
 USB Device (Type B)
Recipe Compounding
 HPLC Buffer Solution Preparation
 mol Conversion Function
 Sample (Concentration) Preparation
 Inspection Support Function
 Clock-CAL
Automatic Doors
Touchless Sensors
Adjustable Windbreak Plate -
Ionizer
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
 Sensitivity Stability Against Temperature Range ±2ppm / ℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W215 × D411 × H346 mm (power supply unit included) Approx. W215 × D367 × H346 mm
Weight Approx. 9.7kg Approx. 8.6kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 1.5A
Power Supply Input (AC Adapter) AC100-240V 480mA 50/60Hz (*7)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Measurement conditions of W-AD series (0.01 mg models only) are as follows.
 - Set the adjustable windbreak plate in the lowest position
 - With a shield plate configured around the pan

*3 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
   The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 Depending on the attached AC adapter.

W Series Analytical Balance

Model AP225W AP135W AP125WD AP225WD AP124W AP224W AP324W
Capacity 220g 135g 120g / 52g 220g / 102g 120g 220g 320g
Minimum Display 0.01mg 0.1mg / 0.01mg 0.1mg
Calibration Weight Built-in (*1)
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 220.00090g
(200g)
45 to 135.00090g
(100g)
45 to 120.00090g
(100g)
95 to 220.00090g
(200g)
45 to 120.0090g
(100g)
95 to 220.0090g
(200g)
95 to 320.0090g
(300g)
 Repeatability (at weighing capacity)
(*3)
0.015mg (up  to 20g)
0.03mg (up to 100g)
0.05mg (up to 200g)
0.05mg 0.1mg / 0.02mg 0.1mg / 0.05mg 0.1mg 0.15mg
Repeatability
(for low loads) (*2, *3)
0.01mg (for 5 g load) 0.1mg
(for 5g load)
0.1mg
(for 10g load)
0.1mg
(for 20g load)
 Minimum Weight (*2, *3) 20mg 200mg
Linearity (*3) ±0.1mg ±0.2mg / ±0.05mg ±0.2mg / ±0.1mg ±0.2mg ±0.3mg
 Response Time for Trace Measurements (*4, *5) 2 sec.
 Response Time (*5) 8 sec. 2 sec. / 8 sec. 2 sec.
 USB Host (Type A)
USB Device (Type B)
Recipe Compounding
HPLC Buffer Solution Preparation
mol Conversion Function
Sample (Concentration) Preparation
Inspection Support Function
Clock-CAL
Internal Windbreak Plate ○ (optional)
Ionizer (optional)
Operational Temperature / Humidity Range 5 to 40℃ at 20 to 85% RH (*6)
Sensitivity Stability Against Temperature Range ±2ppm/℃ (10~30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W213 × D411 × H345 mm (power supply unit included) Approx. W213 × D367 × H345 mm
Weight Approx. 7.9 kg Approx. 7.0 kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 1.0A
Power Supply Input (AC Adapter) AC100 - 240 V / 320mA / 50/60Hz
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB host (Type A)  USB device (Type B)  Ionizer

*1 Minimum display 0.01 mg models provide two internal weights as standard.

*2 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
   The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*3 The value is the result of a test with the internal windbreak plate.

*4 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*5 The response time value is typical.

*6 Non-condensing.

*7 Depending on the attached AC adapter.

X Series / Y Series Analytical Balance

Series X Series Y Series
Model AP324X AP224X AP124X AP324Y AP224Y AP124Y
Capacity 320g 220g 120g 320g 220g 120g
Minimum Display 0.1mg
Calibration Weight Built-in None
External Calibration Weight Range for Span Calibration (Recommended weight value)
95 to 320.0090g
(300g)
95 to 220.0090g
(200g)
45 to 120.0090g
(100g)
95 to 320.0090g
(300g)
95 to 220.0090g
(200g)
45 to 120.0090g
(100g)
Repeatability (at weighing capacity) 0.15mg 0.1mg 0.15mg 0.1mg
Repeatability
(for low loads) (*1)
0.1mg (for 20g load) 0.1mg (for 10g load) 0.1mg (for 5g load) 0.1mg (for 20g load) 0.1mg (for 10g load) 0.1mg (for 5g load)
Minimum Weight (*1) 200mg
Linearity ±0.3mg ±0.2mg ±0.3mg ±0.2mg
 Response Time for Trace Measurements (*2, *3) 2 sec.
 Response Time (*3) 2 sec.
 USB Host (Type A)
USB Device (Type B)
Recipe Compounding
HPLC Buffer Solution Preparation
mol Conversion Function
Sample (Concentration) Preparation
Inspection Support Function
Clock-CAL
Ionizer
Operational Temperature / Humidity Range 5~40℃ 20 to 85% (*4)
Sensitivity Stability Against Temperature Range ±2ppm/℃ (10 to 30℃)
Pan Size Approx. 91 mm dia.
Body Dimensions Approx. W213 × D367 × H345 mm
Weight Approx. 7.0 kg Approx. 6.5 kg
Display OEL display (dot matrix)
Rated Electric Power Supply DC12V 1.0A
Power Supply Input (AC Adapter) AC100 - 240 V / 320mA / 50/60Hz (*5)
Input / Output Terminal RS-232C (D-Sub 9P plug)  USB device (Type B)  Ionizer RS-232C (D-Sub 9P plug)  USB device (Type B) 

*1 Be compliant with USP Chapter 41. This is the tested value by the weight of the balance's capacity of 5 % (or 5 grams' weight)
   The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.

*2 The response time for displaying 90 % of the added sample amount value in trace measurements (from 1 mg).

*3 The response time value is typical.

*4 Non-condensing.

*5 Depending on the attached AC adapter.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3525,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3525,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}