TOC-4200 - Specs
On-line TOC Analyzer
TOC-4200
Measurement items |
TC, NPOC |
---|---|
Data storage function |
Store 20,000 past on-line measured values (equivalent to one year's data at 30-minute measuring cycle). |
Dimensions |
W550 x D383 x H1240 mm (excluding protrusions) |
Construction |
Indoor wall mounting (optional stand available) |
Ambient temperature |
Within 1 to 40℃ |
Power supply |
AC 100 to 240 V ±10%, 10 A, 50/60 Hz |
Carrier gas |
Pressurized air, high-purity air, oxygen*4; supply pressure: 250 to 300 kPa |
Communication function |
Compatible with RS-485, RS-232C (option), and Modbus |
Contact input |
Start calibration, stop measurement, start on-line measurement, reset alarm, measurement start flow line 1 to 6, halt measurement (output contact capacity: 1.4 V DC min., 10 mA min.) |
Contact output signal |
Alarm signals |
Analog output |
Selectable between 4 – 20 mA DC or 0 – 16 mA DC (isolated outputs) |
Printer |
Thermal printer, 40 characters per line; 110 mm chart width (optional) |
Recorder |
Recording width 100 mm, 6-dot recorder (optional) |
Data storage device |
Store measured values or measurement conditions to USB memory*3. |
EPA TOC removal rate calculation function |
TOC removal rate calculation according to the United States EPA regulations (Part IV 40 CFR Part 9, 141 and 142, 1998) (when a 2- or more-stream switching option is provided) |
Measurement principle |
680℃ combustion catalytic oxidation - NDIR detection method |
Load calculation function |
Using the analog input board (option) permits load calculations. (Non-isolated analog inputs; load resistance: 100 ohms) |
Conversion function |
Conversion with a linear expression using required indices (Conversion expression must be determined in advance.) |
Display, operation |
Color LCD with touch screen |
Stream switching function |
Streams 1 to 6 can be switched (with the relevant option) |
Automatic calibration function |
Automatic calibration using standard solution (Dilution water is used as the zero standard solution for zero-calibration.) |
Sample dilution function |
Dilution within syringe; dilution factor 2 to 50 |
IC removal method |
Acidify/sparge within syringe |
Sample injection method |
Syringe pump/sliding injection port |
Measurement cycle |
4 minutes min. (NPOC, residual IC 2 % max.)*2 |
Zero stability |
Within ±2 % f.s. /day (temperature variation within 5℃)*1 |
Span stability |
Within ±2 % f.s. /day (temperature variation within 5℃)*1 |
Repeatability |
Within ±2 % f.s.*1 |
Measurement range |
From 0 - 5 to 0 - 1,000 mgC/L f.s. (0 to 20,000 mgC/L f.s. with dilution function) |
Weight |
Approx. 70 kg (154 lbs.) |
*1) With automatic settings
*2) When measured sample is potassium hydrogen phthalate solution and operations to prepare for the next measurement are performed.
*3) If TN option is installed, oxygen and high-purity nitorogen cannot be used.