NOA-7100 - Specs
Transportable NOx-O2 Gas Analyzer
Standard type
Two standard types are available
Type 1
Measurement components |
NOx, O2 |
---|---|
Wireless data transmission |
Equipped |
Sample gas conditions (*5) |
[ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa; |
External drain separator |
Fitted |
Weight (analyzer) |
Approx. 16 kg |
Dimensions |
W260 x H452 x D420 mm (does not include protrusions) |
Power Requirements |
Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz, |
Permissible ambient humidity (*4). |
Relative humidity 90%RH or less |
Permissible ambient temperature (*4) |
5-40℃; must not be exposed to direct sunlight or radiant heat |
Data output to external media (*3) |
Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection. |
Transmission Output |
0-1VDC, 3 channels, insulated output (individual channels not insulated) |
Measurement range |
NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/4000 ppm |
Display |
Liquid crystal display, able to display 3 items at a time |
Gas sample amount |
Approx. 2 L/min |
Response time (*2) |
NOx: Approx. 20/30/60 sec (Switching is configurable.) |
Linearity |
Within ±2% of full scale |
Span Drift |
Within ±1.5% of full scale per day |
Zero Drift |
Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day) |
Repeatability |
Within ±0.5% of full scale |
Measurement principle |
NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method |
External Signal Input |
0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved) |
Type 2
Measurement components |
NOx, NO, NO2 (*1) |
---|---|
Wireless data transmission |
Equipped |
Sample gas conditions (*5) |
[ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa; |
External drain separator |
Not fitted |
Weight (analyzer) |
Approx. 16 kg |
Dimensions |
W260 x H452 x D420 mm (does not include protrusions) |
Power Requirements |
Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz, |
Permissible ambient humidity (*4). |
Relative humidity 90%RH or less |
Permissible ambient temperature (*4) |
5-40℃; must not be exposed to direct sunlight or radiant heat |
Data output to external media (*3) |
Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection. |
Transmission Output |
0-1VDC, 3 channels, insulated output (individual channels not insulated) |
Measurement range |
NOx, NO, NO2: 0 - 100/ 250/ 500/ 1000/ 2500/ 4000 ppm |
Display |
Liquid crystal display, able to display 3 items at a time |
Gas sample amount |
100 mL/min |
Response time (*2) |
NOx/NO: Within 90 sec |
Linearity |
Within ±2% of full scale |
Span Drift |
Within ±1.5% of full scale per day |
Zero Drift |
Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day) |
Repeatability |
Within ±0.5% of full scale |
Measurement principle |
NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method |
External Signal Input |
0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved) |
Special specifications
In addition to the standard model, it is possible to specify constituents, and if the small flow rate set, communication set, or NOx-NO switching solenoid valve set is provided.
Measurement components |
1 constituent (NOx) or 2 constituents (NOx, O2) (*5) O2 cannot be measured if small flow rate set is selected. See small flow rate set page. NO and NO2 can be measured as well as NOx if NOx-NO switching solenoid valve set is selected. See NOx-NO switching solenoid valve set page. |
Measurement range |
NOx: 0-25/50/100/250/500/1000/2500/4000 ppm O2: 0-5/10/25vol% NOx measurement range differs if small flow rate set is selected. See small flow rate set page. |
Small flow rate set : Not fitted
Response time (*2) |
NOx: Approx. 20/30/60 sec (Switching is configurable.) |
---|---|
Gas sample amount |
Approx. 2 L/min |
Repeatability |
Within ±0.5% of full scale |
Span Drift |
Within ±1.5% of full scale |
Measurement components |
NOx or NOx/O2 |
Measurement range |
NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/ 4000 ppm |
Small flow rate set : Fitted
Response time (*2) |
NOx/NO: Within 90 sec |
---|---|
Gas sample amount |
100 mL/min |
Repeatability |
Within ±0.5% of full scale |
Span Drift |
Within ±1.5% of full scale |
Measurement components |
NOx |
Measurement range |
NOx: 0-100/ 250/ 500/ 1000/ 2500/ 4000 ppm |
Communication set : Not fitted
Wireless data transmission |
Not equipped |
---|---|
Data output to external media (*3) |
Not equipped |
Communication set : Fitted
Wireless data transmission |
Equipped |
---|---|
Data output to external media (*3) |
Measurement data stored in the internal memory of the analyzer can be output to USB flash drives in CSV file format; can also be output to computer via Wi-Fi or LAN cable connection. |
This special accessory switches the flow path inside the analyzer between a path that flows through the converter (NOx flow path) and one that does not flow through the converter (NO path). The switch cycle is selectable from auto (configurable from 1 to 99 minutes), manual, and fixed. If there is significant variation over time in concentrations of NOx, NO, or NO2 relative to the switching cycle, NO2 concentration measurement accuracy diminishes.
Basic specifications are as follows.
Basic specifications
Measurement principle | NOx: Atmospheric pressure chemiluminescence method O2: Limiting current type zirconia method |
Repeatability | Within ±0.5% of full scale |
Zero drift | Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day) |
Span drift | Within ±1.5% of full scale |
Linearity | Within ±2% of full scale |
The following specifications are the same as for the standard model.
Display, transmission output, permissible ambient temperature, permissible ambient humidity, power requirements, dimensions, weight (body), sample gas conditions, external signal input.
*1: |
NO2 is measured by calculation (NOx-NO). |
*2: |
Time that indicates 90% response from gas flowing from sample gas inlet (flowing gas switching time + electrical system response time). With Type 2, this will be the response when NOx-NO switching is fixed. |
*3: |
Check the Shimadzu website for smart devices and USB flash drives for which operation has been confirmed. |
*4: |
Permissible ambient temperature and humidity are indicated by the shaded part of the figure. |
*5: |
Sample gas conditions and coexisting gases are summarized below. Please contact your Shimadzu representative if the conditions are not satisfied. |