NOA-7100 - Specs

Transportable NOx-O2 Gas Analyzer

Standard type

Two standard types are available

Type 1

Measurement components

NOx, O2

Wireless data transmission

Equipped

Sample gas conditions (*5)

[ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa;
Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less

External drain separator

Fitted

Weight (analyzer)

Approx. 16 kg

Dimensions

W260 x H452 x D420 mm (does not include protrusions)

Power Requirements

Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz,
100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)

Permissible ambient humidity (*4).

Relative humidity 90%RH or less

Permissible ambient temperature (*4)

5-40℃; must not be exposed to direct sunlight or radiant heat

Data output to external media (*3)

Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection.

Transmission Output

0-1VDC, 3 channels, insulated output (individual channels not insulated)

Measurement range

NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/4000 ppm
O2: 0 - 5/ 10/ 25 vol%

Display

Liquid crystal display, able to display 3 items at a time

Gas sample amount

Approx. 2 L/min

Response time (*2)

NOx: Approx. 20/30/60 sec (Switching is configurable.)
O2: Approx. 30/60 sec (Switching is configurable.)

Linearity

Within ±2% of full scale

Span Drift

Within ±1.5% of full scale per day

Zero Drift

Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)

Repeatability

Within ±0.5% of full scale

Measurement principle

NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method

External Signal Input

0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)

Read more
Show Less

Type 2

Measurement components

NOx, NO, NO2 (*1)

Wireless data transmission

Equipped

Sample gas conditions (*5)

[ At NOA-7100 inlet port ] Pressure: -980 to + 980 Pa;
Temperature: Ambient temperature; Dust: 0.1 g/Nm3 or less

External drain separator

Not fitted

Weight (analyzer)

Approx. 16 kg

Dimensions

W260 x H452 x D420 mm (does not include protrusions)

Power Requirements

Choose from 110, 115, 120, 127, 220, 230, 240 VAC 50-60 Hz,
100 V series: 410 VA, 200 V series: 700 VA (steady 200 VA)

Permissible ambient humidity (*4).

Relative humidity 90%RH or less

Permissible ambient temperature (*4)

5-40℃; must not be exposed to direct sunlight or radiant heat

Data output to external media (*3)

Measurement data stored in the internal memory of the analyzer can be written to USB flash drives in CSV file format; can also be output to a computer via Wi-Fi or LAN cable connection.

Transmission Output

0-1VDC, 3 channels, insulated output (individual channels not insulated)

Measurement range

NOx, NO, NO2: 0 - 100/ 250/ 500/ 1000/ 2500/ 4000 ppm

Display

Liquid crystal display, able to display 3 items at a time

Gas sample amount

100 mL/min

Response time (*2)

NOx/NO: Within 90 sec

Linearity

Within ±2% of full scale

Span Drift

Within ±1.5% of full scale per day

Zero Drift

Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)

Repeatability

Within ±0.5% of full scale

Measurement principle

NOx: Atmospheric pressure chemiluminescence method; O2: Limiting current zirconia type method

External Signal Input

0-1VDC, 1 channel, Signal input from an external analyzer, display (Data can be saved)

Read more
Show Less

Special specifications

In addition to the standard model, it is possible to specify constituents, and if the small flow rate set, communication set, or NOx-NO switching solenoid valve set is provided.

Measurement
components
1 constituent (NOx) or 2 constituents (NOx, O2) (*5)
O2 cannot be measured if small flow rate set is selected. See small flow rate set page.
NO and NO2 can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.
See NOx-NO switching solenoid valve set page.
Measurement
range
NOx: 0-25/50/100/250/500/1000/2500/4000 ppm
O2: 0-5/10/25vol%
NOx measurement range differs if small flow rate set is selected.
See small flow rate set page.

Small flow rate set : Not fitted

Response time (*2)

NOx: Approx. 20/30/60 sec (Switching is configurable.)
O2: Approx. 30/60 sec (Switching is configurable.)

Gas sample amount

Approx. 2 L/min

Repeatability

Within ±0.5% of full scale

Span Drift

Within ±1.5% of full scale

Measurement components

NOx or NOx/O2
NO and NO2 (*1) can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.

Measurement range

NOx: 0 - 25/ 50/ 100/ 250/ 500/ 1000/ 2500/ 4000 ppm
O2: 0-5/ 10/ 25 vol%

Read more
Show Less

Small flow rate set : Fitted

Response time (*2)

NOx/NO: Within 90 sec

Gas sample amount

100 mL/min

Repeatability

Within ±0.5% of full scale

Span Drift

Within ±1.5% of full scale

Measurement components

NOx
NO and NO2 (*1) can be measured as well as NOx if NOx-NO switching solenoid valve set is selected.

Measurement range

NOx: 0-100/ 250/ 500/ 1000/ 2500/ 4000 ppm

Read more
Show Less

Communication set : Not fitted

Wireless data transmission

Not equipped

Data output to external media (*3)

Not equipped

Communication set : Fitted

Wireless data transmission

Equipped

Data output to external media (*3)

Measurement data stored in the internal memory of the analyzer can be output to USB flash drives in CSV file format; can also be output to computer via Wi-Fi or LAN cable connection.

NOx-NO switching solenoid valve set : Not fitted

Measurement components

NOx

NOx-NO switching solenoid valve set : Fitted

Measurement components

NOx, NO, NO2 (*1)

This special accessory switches the flow path inside the analyzer between a path that flows through the converter (NOx flow path) and one that does not flow through the converter (NO path). The switch cycle is selectable from auto (configurable from 1 to 99 minutes), manual, and fixed. If there is significant variation over time in concentrations of NOx, NO, or NO2 relative to the switching cycle, NO2 concentration measurement accuracy diminishes.

Basic specifications are as follows.

Basic specifications

Measurement principle NOx: Atmospheric pressure chemiluminescence method
O2: Limiting current type zirconia method
Repeatability Within ±0.5% of full scale
Zero drift Within ±1% of full scale per day (25 ppm range: within ±1.5% of full scale per day)
Span drift Within ±1.5% of full scale
Linearity Within ±2% of full scale

The following specifications are the same as for the standard model.
Display, transmission output, permissible ambient temperature, permissible ambient humidity, power requirements, dimensions, weight (body), sample gas conditions, external signal input.

*1:

NO2 is measured by calculation (NOx-NO).

*2:

Time that indicates 90% response from gas flowing from sample gas inlet (flowing gas switching time + electrical system response time). With Type 2, this will be the response when NOx-NO switching is fixed.

*3:

Check the Shimadzu website for smart devices and USB flash drives for which operation has been confirmed.

*4:

Permissible ambient temperature and humidity are indicated by the shaded part of the figure.

*5:

Sample gas conditions and coexisting gases are summarized below. Please contact your Shimadzu representative if the conditions are not satisfied.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3498,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3498,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}