PPSQ-51A/53A - Specs
Protein Sequencer
Main Unit | PPSQ-51A | PPSQ-53A |
Reaction method | Edman degradation | |
Reaction time | 46.5 min/cycle | 48 min/cycle |
Number of reactors | 1 | 3 |
Sample immobilization method | Glass fiber disk (8 mm dia.) or PVDF membrane | |
Reactor temperature control range | 10 °C above room temperature to 60 °C | |
Converter temperature control range | 10 °C above room temperature to 70 °C | |
Number of reagents/solvents | 7 | |
Reagent/solvent delivery method | Nitrogen gas pressure | |
Dimensions | W510 x D500 x H540 mm | |
Weight | 43 kg | 45 kg |
Other Items Provided by User
Installation space | Desktop: W1,600 x D600 x H800 mm min. |
Weight: Approx. 100 kg | |
Power supply | 120-230 VAC, 50/60 Hz, 1,500 VA max. |
Nitrogen gas | Purity: 99.9999% min. |
A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories. | |
Exhaust equipment | In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory. |
Note: The required installation space and power supply may change according to the PC and display used.
Installation Example