RF-20A/RF-20Axs - Specs
Fluorescence Detector
RF-20A
Fluorescence Detector
Light source |
Xenon lamp |
---|---|
Wavelength range |
0, 200 nm to 650 nm |
Spectral bandwidth |
20nm |
Wavelength accuracy |
±2 nm |
Wavelength Reproducibility |
±0.2 nm |
S/N |
Water Raman peak S/N 1200 min. |
Cell (capacity, pressure resistance, material) |
12 µL; 2 MPa (approx. 20 kgf/cm2); SUS316L, PTFE (fluororesin), quartz |
RF-20Axs
Fluorescence Detector
Light source |
Xenon lamp, Low-pressure mercury lamp (to check wavelength accuracy) |
---|---|
Wavelength range |
0, 200 nm to 750 nm |
Spectral bandwidth |
20nm |
Wavelength accuracy |
±2 nm |
Wavelength Reproducibility |
±0.2 nm |
S/N |
Water Raman peak S/N 2000min. |
Cell (capacity, pressure resistance, material) |
12 µL; 2 MPa (approx. 20 kgf/cm2); SUS316L, PTFE (fluororesin), quartz |
Cell Temperature Input Range |
4°C |
Cell Temperature Control Range |
(Room temperature - 10°C) to 40°C |
What is a fluorescence detector?
A fluorescence detector excites the sample with excitation light and breaks up the emitted fluorescence light with a fluorescence monochromator. It extracts the required fluorescence wavelengths and measures the intensity with a photomultiplier.