0

    Strict Manufacturing Uniformity

    Comparison of reproducibility between three batches of silica-based materials

    Balance of Number of Theoretical Plates and Column Pressure

    Shim-pack VP series achieves a higher number of theoretical plates while maintaining low pressure. It provides superior performance shown by NTP per 1MPa.

    NTP per 1MPa
    Column Size: 4.6 x 150 mm, Mobile Phase: Methanol/Water=70/30,
    Flow Rate: 1.0 mL/min., Sample: Naphthalene

    Excellent Peak Shape

    Packing materials with less metal impurities achieve excellent peak shape of coordination compounds.

    Completed end-capping achieves excellent peak shape of basic compounds.

    Smooth Transfer of Methods from High-Speed Analysis to Conventional Analysis

    By using the Shim-pack VP series of conventional columns in combination with the Shim-pack XR series of high-speed columns, which offer the equivalent separation performance, it is possible to perform method development via high-speed analysis, and then smoothly transfer to conventional analysis. Utilizing these columns with Nexera Method Scouting provides strong support for method development.

    Conditions
    Column Top:
    Shim-pack VP-ODS
    (150 mmL. × 4.6 mmI.D., 5 μm)
    (P/N: 228-34937-91)
        Bottom:
    Shim-pack XR-ODS
    (50 mmL. × 3.0 mmI.D., 2.2 μm)
    (P/N: 228-41606-92)
    Mobile Phase A) 0.1 % formic acid
    B) Acetonitrile
    A/B = 8/2 (v/v)
    Flow Rate Top: 1.0 mL/min
    Bottom: 1.0 mL/min
     
     
     

    Relative Retention Times

    Peak Shim-pack VP-C8 Shim-pack XR-C8
      1. Sulfadiamine 1.000 1.000
      2. Sulfamerazine 1.252 1.238
     3. Sulfadimidine 1.537 1.509
     4. Sulfamethoxazole 2.809 2.699
     5. Sulfaquinoxaline 5.668 5.432

    Analysis of Fat-Soluble Vitamin

    Conditions
    Column Top:
    Shim-pack VP-C8
    (150 mmL. × 4.6 mmI.D., 5 μm)
    (P/N: 228-59927-91)
      Bottom:
    Shim-pack XR-C8
    (50 mmL. × 3.0 mmI.D., 2.2 μm)
    (P/N: 228-59902-92)
    Mobile Phase Methanol
    Flow Rate Top: 1.0 mL/min
    Bottom: 1.0 mL/min
     

    Relative Retention Times

    Peak Shim-pack VP-C8 Shim-pack XR-C8
    1. Retinol 1.000 1.000
    2. α±−tocopherol 1.509 1.531


     

     
     

    Analysis of Fat-Soluble Vitamin

    Conditions
    Column Top:
    Shim-pack VP-C8
    (150 mmL. × 4.6 mmI.D., 5 μm)
    (P/N: 228-59927-91)
      Bottom:
    Shim-pack XR-C8
    (50 mmL. × 3.0 mmI.D., 2.2 μm)
    (P/N: 228-59902-92)
    Mobile Phase Methanol
    Flow Rate Top: 1.0 mL/min
    Bottom: 1.0 mL/min

     

    Relative Retention Times

    Peak Shim-pack VP-C8 Shim-pack XR-C8
    1. Retinol 1.000 1.000
    2. α±−tocopherol 1.509 1.531
     

    Batch Analysis of Non-Steroidal Anti-Inflammatory Drugs (NSAIDs)

    Conditions
    Column Top:
    Shim-pack VP-Phenyl
    (150 mmL. × 4.6 mmI.D., 5 μm)
    (P/N: 228-59928-91)
      Bottom:
    Shim-pack XR-Phenyl
    (50 mmL. × 3.0 mmI.D., 2.2 μm)
    (P/N: 228-59904-92)
    Mobile Phase A) 20 mmol/L Phosphoric acid (Na) buffer (pH 2.5)
    B) Acetonitrile
    A/B = 3/7 (v/v)
    Flow Rate Top: 1.0 mL/min
    Bottom: 1.0 mL/min

     

    Relative Retention Times

    Peak Shim-pack VP-Phenyl Shim-pack XR-Phenyl
    1. Piroxicam 1.000 1.000
    2. Ketoprofen 1.480 1.483
    3. Diflunisal 2.233 2.172
    4. Diclofenac 2.949 2.916
    5. Indomethacin 3.263 3.262
    6. Ibuprofen 3.470 3.501

    Specifications

      Shim-pack VP-ODS Shim-pack VP-C8 Shim-pack VP-Phenyl
     Particle Size (µm)  5  5  5
     Pore Size (nm)  12  12  12
     Surface Area (m2/g)  410  410  410
     Carbon Loading  20%  12.5%  12.3%
     Pressure Tolerance (MPa)  Approx. 20  Approx. 20  Approx. 20
     Pore Volume (mL/g)  1.25  1.25  1.25
     End-capping  Yes  Yes  Yes
     Bonding Type  Monomeric  Monomeric  Monomeric
     pH range  2-7.5  2-7.5  2-7.5
     USP Code  L1  L7  L11

    * Appearance and specifications are subject to change without notice.