TCE-N300A - Specs

Compact Thermostatic Chamber for Autograph Precision Universal Testing Machines

Main Unit - Standard Height

Temperature Usage Range *1 *2

-70 to +300 ℃

Heating system

Heater

Cooling System *3

Liquid Nitrogen Gas Spray

External Dimensions - Width

Caster Type: 680 mm; Rail Type: 410 mm

External Dimensions - Depth

Caster Type: 860 mm; Rail Type: 1280 mm

External Dimensions - Height

Caster Type: 1405 to 1815 mm

Rail Type: 1300 to 1730 mm

Chamber Inner Dimension - Width

300 mm

Chamber Inner Dimension - Depth

300 mm

Chamber Inner Dimension - Height

600 mm

Weight

Caster Type: 80 kg

Rail Type: 70 kg

Power requirement

Single-phase 200 to 230 V, 50/60 Hz

Power Capacity

2.7 kVA for 200 V; 3.5 kVA for 230 V

Temperature Increase Rate *4

25 minutes from room temperature to 300 ℃

Temperature Decrease Rate *4

25 minutes from room temperature to -70 ℃

Temperature Stability *4 *5

Within ±2 ℃

Temperature distribution

Minimum to maximum for three measurement locations is within 3 ℃

Temperature Distribution Measurement Points *6

Three points: The center of the thermostatic chamber, and 110 mm below and above the center

*1: The indicated temperature for the thermostatic chamber is measured by a resistance temperature detector (Pt 100) situated within the thermostatic chamber at the back (in front of the fan) at center height.

*2: There could be a difference between the indicated temperature for the thermostatic chamber and the temperature of the actual sample. In this case, measure the temperature of the actual sample, and change the temperature settings of the thermostatic chamber to obtain the target temperature.

*3: Use a cooling gas cylinder that satisfies the specifications noted below.

Gas: Liquid nitrogen (LN2)

Container: Pressurized container (with siphon tube and safety valve)

Gas pressure: 0.05 to 0.1 MPa

*4: This is the rate for an empty chamber when the vertical pull rod insertion port is packed with an insulating material. In addition, this rate is when the power voltage is maintained at 200 V.

*5: A thermostatic chamber control thermometer is used for measurements within 30 minutes after the temperature is reached.

*6: A thermometer inserted in a 30 mm aluminum rod situated within the chamber at the center is used for these measurements.

Read more
Show Less

Main Unit - 125 mm Extension Type

Temperature Usage Range *1 *2

-70 to +300 ℃

Heating system

Heater

Cooling System *3

Liquid Nitrogen Gas Spray

External Dimensions - Width

Caster Type: 680 mm; Rail Type: 410 mm

External Dimensions - Depth

Caster Type: 860 mm; Rail Type: 1280 mm

External Dimensions - Height

Rail Type: 1420 to 1850 mm

Chamber Inner Dimension - Width

300 mm

Chamber Inner Dimension - Depth

300 mm

Chamber Inner Dimension - Height

725 mm

Weight

Rail Type: 76 kg

Power requirement

Single-phase 200 to 230 V, 50/60 Hz

Power Capacity

2.7 kVA for 200 V; 3.5 kVA for 230 V

Temperature Increase Rate *4

25 minutes from room temperature to 300 ℃

Temperature Decrease Rate *4

25 minutes from room temperature to -70 ℃

Temperature Stability *4 *5

Within ±2 ℃

Temperature distribution

Minimum to maximum for three measurement locations is within 3 ℃

Temperature Distribution Measurement Points *6

Three points: The center of the thermostatic chamber, and 132.5 mm below and above the center

*1: The indicated temperature for the thermostatic chamber is measured by a resistance temperature detector (Pt 100) situated within the thermostatic chamber at the back (in front of the fan) at center height.

*2: There could be a difference between the indicated temperature for the thermostatic chamber and the temperature of the actual sample. In this case, measure the temperature of the actual sample, and change the temperature settings of the thermostatic chamber to obtain the target temperature.

*3: Use a cooling gas cylinder that satisfies the specifications noted below.

Gas: Liquid nitrogen (LN2)

Container: Pressurized container (with siphon tube and safety valve)

Gas pressure: 0.05 to 0.1 MPa

*4: This is the rate for an empty chamber when the vertical pull rod insertion port is packed with an insulating material. In addition, this rate is when the power voltage is maintained at 200 V.

*5: A thermostatic chamber control thermometer is used for measurements within 30 minutes after the temperature is reached.

*6: A thermometer inserted in a 30 mm aluminum rod situated within the chamber at the center is used for these measurements.

Read more
Show Less

Main Unit - 250 mm Extension Type

Temperature Usage Range *1 *2

-70 to +300 ℃

Heating system

Heater

Cooling System *3

Liquid Nitrogen Gas Spray

External Dimensions - Width

Caster Type: 680 mm; Rail Type: 410 mm

External Dimensions - Depth

Caster Type: 860 mm; Rail Type: 1280 mm

External Dimensions - Height

Rail Type: 1550 to 1980 mm

Chamber Inner Dimension - Width

300 mm

Chamber Inner Dimension - Depth

300 mm

Chamber Inner Dimension - Height

850 mm

Weight

Rail Type: 82 kg

Power requirement

Single-phase 200 to 230 V, 50/60 Hz

Power Capacity

2.7 kVA for 200 V; 3.5 kVA for 230 V

Temperature Increase Rate *4

25 minutes from room temperature to 300 ℃

Temperature Decrease Rate *4

25 minutes from room temperature to -70 ℃

Temperature Stability *4 *5

Within ±2 ℃

Temperature distribution

Minimum to maximum for three measurement locations is within 3 ℃

Temperature Distribution Measurement Points *6

Three points: The center of the thermostatic chamber, and 155 mm below and above the center

*1: The indicated temperature for the thermostatic chamber is measured by a resistance temperature detector (Pt 100) situated within the thermostatic chamber at the back (in front of the fan) at center height.

*2: There could be a difference between the indicated temperature for the thermostatic chamber and the temperature of the actual sample. In this case, measure the temperature of the actual sample, and change the temperature settings of the thermostatic chamber to obtain the target temperature.

*3: Use a cooling gas cylinder that satisfies the specifications noted below.

Gas: Liquid nitrogen (LN2)

Container: Pressurized container (with siphon tube and safety valve)

Gas pressure: 0.05 to 0.1 MPa

*4: This is the rate for an empty chamber when the vertical pull rod insertion port is packed with an insulating material. In addition, this rate is when the power voltage is maintained at 200 V.

*5: A thermostatic chamber control thermometer is used for measurements within 30 minutes after the temperature is reached.

*6: A thermometer inserted in a 30 mm aluminum rod situated within the chamber at the center is used for these measurements.

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4623,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4623,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}