Autograph AGS-X Series - Specs
AUTOGRAPH Table-TOP Precision Universal Tester
AGS-10kNXD
Specifications Table-top
Max. Load Capacity |
10kN |
---|---|
Crosshead Position Detection:Display Method |
Digital display (display resolution: 0.001 mm) |
Power Requirements |
Supply voltage fluctuations within ±10% of the set value. |
Power Requirements |
Single phase 100/120/220/240 V AC |
Weight |
85kg |
Dimensions |
W653×D520× |
Accessories |
Load cell (with CAL cable), Power cable (2.5 m),turning rod, cable clamps, instruction manual |
Standard Functions |
• Analog recorder (option) output |
Standard Functions |
• External analog input (2 channels) *4 |
Standard Functions |
• Automatic reading of load cell characteristic values |
Sampling Speed |
1000Hz max. *4 |
Crosshead Position Detection:Positional Accuracy |
±0.1% indicated value or ±0.01 mm, whichever is larger |
Crosshead Position Detection:Measurement Method |
Optical encoder |
Lineup of Models Differing byLoad Cell Capacity |
1N /2N /5N /10N /20N /50N |
Effective Test Width |
425mm |
Crosshead – Table Distance(Tensile stroke) *3 |
1200mm |
Crosshead Speed and Permitted Test Force |
To load cell capacity across entire speed range |
Crosshead Speed Accuracy*2 |
±0.1 |
Crosshead : Max. Return Speed |
1500mm/min |
Crosshead : Speed Range |
0.001 to 1000 mm/min (stepless) |
Measurement : Calibration |
Automatic test force calibration: select tensile, compression, or |
Measurement : Accuracy : Standard-Precision Type (1/500, ±1%) |
Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
Force:Accuracy : High-Precision Type (1/500, ±0.5%) |
Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) |
Loading Method |
Direct high-precision constant speed strain control method via non-backlash ball-screw drive |
Operating Environment |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-50kNXD
Specifications Table-top
Max. Load Capacity |
50kN |
---|---|
Crosshead Position Detection : Display Method |
Digital display (display resolution: 0.001 mm) |
Power Requirements |
Supply voltage fluctuations within ±10% of the set value. |
Power Requirements |
Single phase AC200-230V 50/60Hz 4.0kVA |
Weight |
260kg |
Dimensions |
W718×D641× |
Accessories |
Load Cell (with CAL cable), Power cable (5 m), turning rod, cable clamps, instruction manual |
Standard Functions |
• Analog recorder (option) output |
Standard Functions |
• External analog input (2 channels) *4 |
Standard Functions |
• Automatic reading of load cell characteristic values |
Sampling Speed |
1000Hz max. *4 |
Crosshead Position Detection : Positional Accuracy |
±0.1% indicated value or ±0.01 mm, whichever is larger |
Crosshead Position Detection : Measurement Method |
Optical encoder |
Lineup of Models Differing byLoad Cell Capacity |
50kN |
Effective Test Width |
425mm |
Crosshead – Table Distance(Tensile stroke) *3 |
1210mm |
Crosshead Speed and Permitted Test Force |
To load cell capacity across entire speed range |
Crosshead Speed Accuracy*2 |
±0.1 |
Crosshead : Max. Return Speed |
1100mm/min |
Crosshead : Speed Range |
0.001 to 800 mm/min (stepless) |
Measurement : Calibration |
Automatic test force calibration: select tensile, compression, or |
Measurement : Accuracy : Standard-Precision Type (1/500, ±1%) |
Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
Force : Accuracy : High-Precision Type (1/500, ±0.5%) |
Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) |
Loading Method |
Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive |
Operating Environment |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-100kNX
Specifications Floor Type
Capacity |
100kN |
---|---|
Crosshead Position Detection : Display Method |
Digital display (display resolution: 0.001 mm) |
Dimensions |
W945×D725×H2164mm |
Accessories |
Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual |
Standard Functions |
• External digital input (2 channels) *4 |
Standard Functions |
• External analog input (2 channels) *4 |
Standard Functions |
• Automatic reading of load cell characteristic values |
Test Method Files |
40 files (PC link: 20 files, standalone controller: 20 files) |
Sampling Speed |
1000Hz max. *4 |
Crosshead Position Detection : Positional Accuracy |
±0.1% indicated value or ±0.01 mm, whichever is larger |
Crosshead Position Detection : Measurement Method |
Optical encoder |
Loading Method |
Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive |
Effective Test Width |
600mm |
Crosshead - Table Distance (Tensile stroke) *3 |
1255mm (745mm, MWG) |
Crosshead Speed and Permitted Test Force |
To load cell capacity across entire speed range |
Crosshead Speed Accuracy *2 |
±0.1% |
Crosshead : Max. Return Speed |
1100mm/min |
Crosshead : Speed Range |
0.001 to 800 mm/min (stepless) |
Force Measurement : Calibration |
Automatic test force calibration: select tensile, compression, or tensile and compression |
Force Measurement : Accuracy : Standard-Precision Type(1/500, ±1%) |
Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
Force Measurement : Accuracy : High-Precision Type(1/500, ±0.5%) |
Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) |
Weight |
525kg |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-300kNX
Specifications Floor Type
Capacity |
300kN |
---|---|
Crosshead Position Detection : Display Method |
Digital display (display resolution: 0.001 mm) |
Dimensions |
W945×D725×H2414mm |
Accessories |
Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual |
Standard Functions |
• External digital input (2 channels) *4 |
Standard Functions |
• External analog input (2 channels) *4 |
Standard Functions |
• Automatic reading of load cell characteristic values |
Test Method Files |
40 files (PC link: 20 files, standalone controller: 20 files) |
Sampling Speed |
1000Hz max. *4 |
Crosshead Position Detection : Positional Accuracy |
±0.1% indicated value or ±0.01 mm, whichever is larger |
Crosshead Position Detection : Measurement Method |
Optical encoder |
Loading Method |
Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive |
Effective Test Width |
600mm |
Crosshead - Table Distance (Tensile stroke) *3 |
1475mm (635mm, MWG) |
Crosshead Speed and Permitted Test Force |
0.001 to 200kN:0.001 to 500mm/min(entire speed range) 200kN to 300kN:0.001 to 400mm/min |
Crosshead Speed Accuracy *2 |
±0.1% |
Crosshead : Max. Return Speed |
550mm/min |
Crosshead : Speed Range |
0.001 to 500 mm/min (stepless) |
Force Measurement : Calibration |
Automatic test force calibration: select tensile, compression, or tensile and compression |
Force Measurement : Accuracy : Standard-Precision Type(1/500, ±1%) |
Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
Force Measurement : Accuracy : High-Precision Type(1/500, ±0.5%) |
Within ±0.5% indicated test force (at 1/250 to 1/1 load cell rating) |
Weight |
675kg |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-100kNX
Specifications Floor Type
Power Requirements : 400V model |
Three phases AC380-440V |
---|---|
Operating Environment : 400V model |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
Power Requirements : 200V model |
A three phase AC200-230V |
Operating Environment : 200V model |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
Power Requirements : 400V model |
Three phases |
Operating Environment : 400V model |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
Power Requirements : 200V model |
A three phase |
Operating Environment : 200V model |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AUTOGRAPH Table-Top Precision Universal Tester
Load capacity |
1 N to 10 kN (11 types) |
---|---|
Test speed |
0.001 to 1,000 mm/min (Stepless) |
Test force accuracy |
Within ±0.5 % of display test force (for 1/1 to 1/500 of load cell capacity) |
AGS-X Series
Load capacity |
1 N to 10 kN (11 types) |
---|---|
Test speed |
0.001 to 1,000 mm/min (Stepless) |
Test force accuracy |
Within ±0.5 % of display test force (for 1/1 to 1/500 of load cell capacity) |
AGS-20kNXD
Specifications Table-top
Max. Load Capacity |
20kN |
---|---|
Crosshead Position Detection:Display Method |
Digital display (display resolution: 0.001 mm) |
Power Requirements |
Supply voltage fluctuations within ±10% of the set value. |
Power Requirements |
Single phase AC200-230V 50/60Hz 4.0kVA |
Weight |
235kg |
Dimensions |
W718×D641× |
Accessories |
Load Cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual |
Standard Functions |
• Analog recorder (option) output |
Standard Functions |
• External analog input (2 channels) *4 |
Standard Functions |
• Automatic reading of load cell characteristic values |
Sampling Speed |
1000Hz max. *4 |
Crosshead Position Detection:Positional Accuracy |
±0.1% indicated value or ±0.01 mm, whichever is larger |
Crosshead Position Detection:Measurement Method |
Optical encoder |
Lineup of Models Differing byLoad Cell Capacity |
20kN |
Effective Test Width |
425mm |
Crosshead – Table Distance(Tensile stroke) *3 |
1250mm |
Crosshead Speed and Permitted Test Force |
To load cell capacity across entire speed range |
Crosshead Speed Accuracy*2 |
±0.1 |
Crosshead : Max. Return Speed |
2200mm/min |
Crosshead : Speed Range |
0.001 to 1600 mm/min (stepless) |
Measurement : Calibration |
Automatic test force calibration: select tensile, compression, or |
Measurement : Accuracy:Standard-Precision Type (1/500, ±1%) |
Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) |
Force:Accuracy : High-Precision Type (1/500, ±0.5%) |
Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) |
Loading Method |
Direct high-precision constant speed strain control method via non-backlash ball-screw drive |
Operating Environment |
Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) |
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation