Main Specifications

X-Ray Generator Output 160 kV   1.2 mA
X-Ray Detector Flat panel detector
Max. Input Resolution Approx. 5.6 million pixels
Max. Sample Size/Weight SRD*1 = 475 mm ϕ300 mm×H 320 mm,10 kg (Including sample setting jig)
SRD = 550 mm φ150 mm×H 320 mm,10 kg (Including sample setting jig)
Max. CT-Scan Region SRD = 475 mm φ85 mm×H 70 mm
SRD = 550 mm φ100 mm×H 80 mm
Spatial Resolution 100 µm (SRD = 550 mm)
CT Data Acquisition Time 12 sec or 5 min
Rated
Output
Main Unit Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
Control Computer Single phase 100 to 240 V AC ± 10 %, 50/60 Hz, 1.0 kVA
Grounding D type grounding (Grounding resistance of 100  max.)
Weight 290 kg
Operating Environment Conditions Ambient temperature: 10 to 30 °C; Ambient humidity: RH of 45 to 80 % max. (No condensation)
External Leakage Dose 1Sv/h or less.

*1 SRD: The source-to-rotation center distance (SRD) is the distance from the X-ray source to the rotation center of the sample.

XSeeker 8000 Image and Layout (Units: mm)

XSeeker 8000 Image and Layout (Units: mm)

Recommended Installation Space: W1950 × D1000 mm
Note: Leave Leave at least 350 mm of space behind the CT system table and 150mm of space left side in order to permit access for maintenance work. 

XSeeker and the Analytical Intelligence logo are trademarks of Shimadzu Corporation or its affiliated companies in Japan and/or other countries.
VGSTUDIO MAX is a trademark of Volume Graphics GmbH.

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":4662,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":4662,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}