    Main Unit PPSQ-51A PPSQ-53A
    Reaction method Edman degradation
    Reaction time 46.5 min/cycle 48 min/cycle
    Number of reactors 1 3
    Sample immobilization method Glass fiber disk (8 mm dia.) or PVDF membrane
    Reactor temperature control range 10 °C above room temperature to 60 °C
    Converter temperature control range 10 °C above room temperature to 70 °C
    Number of reagents/solvents 7
    Reagent/solvent delivery method Nitrogen gas pressure
    Dimensions W510 x D500 x H540 mm
    Weight 43 kg 45 kg

    PPSQ-51A
    Single Reactor


    Single Reactor

    Offers highly cost-effective performance.

    PPSQ-53A
    Triple Reactor
    (Supports consecutive analysis of 3 samples.)

    Triple Reactor

    Using a triple reactor makes it possible to save time and effort, and gives greater freedom in the formulation of analysis programs.

    Standard Configurations

    Isocratic System

    Configuration PPSQ-51A/53A (main unit)
    LC-40D
    SPD-M30A
    CTO-40C

    Gradient System

    Configuration PPSQ-51A/53A (main unit)
    LC-40D (two)
    SPD-M30A
    CTO-40C
    DGU-403

    Other Items Provided by User

    Installation space Desktop: W1,600 x D600 x H800 mm min.
    Weight: Approx. 100 kg
    Power supply 120-230 VAC, 50/60 Hz, 1,500 VA max.
    Nitrogen gas Purity: 99.9999% min.
    A cylinder pressure regulator and gas tubing (10 m) are provided as standard accessories.
    Exhaust equipment In order to provide an exhaust for gases produced by waste liquids, an exhaust tube must either be connected to exhaust equipment or vented outside. An exhaust tube (20 m) is provided as a standard accessory.

    Note: The required installation space and power supply may change according to the PC and display used.

     

    Installation Example

    Isocratic System

    Isocratic System

    Gradient System

    Gradient System