Standard specifications
Turbo molecular pump (Note 1) | TMP-V2304LM | |
---|---|---|
Inlet flange | VG250 / ISO250B | |
Outlet flange | KF40 | |
Cooling method | Water | |
Ultimate pressure (after baking) | 10-7 Pa order | |
Maximum allowable Argon gas flow rate (Note 2, Note 3) | 600 mL/min | |
Maximum allowable inlet pressure (N2 continuous exhaust) |
4.5 Pa | |
Maximum allowable outlet pressure (N2 continuous exhaust) |
200 Pa | |
Pumping speed (Note 4) | N2 | 2100 L/s |
Ar | 2000 L/s | |
Compression ratio | N2 | 1×108 or more |
He | 5×103 | |
H2 | 9×102 | |
Rated speed | 33700 rpm | |
Start-up time | 9 minutes or less | |
Mounting position | In any desired direction | |
Vibration level (by Shimadzu's method) | 0.01 μm or less(single amplitude) | |
Recommended flow rate of purge gas (Note 2) | 30 mL/min | |
Recommended pumping speed of backing pump in case of gas purge | 1500 L/min or more | |
Environment temperature | 10 ~ 40 ℃ | |
Admissible ambient magnetic field | Radial direction | 3 mT |
Axial direction | 15 mT | |
Water | Flow Rate | 3 ~ 4 L/min |
Pressure | 0.2 ~ 0.4 MPa | |
Temperature | 19 ~ 30 degrees C.(above dew point) | |
Display | LED | POWER / STATUS / ROTATION / NET |
Communication | Contact | REMOTE(D-sub 37 pin male, Screw lock size M2.6) |
Input:START / STOP / RESET / LOW SPEED | ||
Output:ROTATION / ACC. / BRAKE / NORMAL /REMOTE / ALARM / WARNING / CONNECTION | ||
Serial | RS-232C / RS-485(D-sub 9 pin male, Screw lock size M2.6) | |
Speed variation | Speed is variable between 25 % and 100 % of the rated speed. (set as 0.1 %) | |
Input electric power | Voltage | Single phase AC 200 ~ 240 V(50 / 60 Hz) |
Maximum power | 1.2 kVA | |
Mass | 56 kg |
- (Note 1)Do not use corrosive gas because there is possibility that the pump breaks down by the product material.
- (Note 2)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
- (Note 3)The maximum allowable flow rate depends on the gas type, the APC valve temperature, the gas purge flow rate and outlet pressure.
- (Note 4)Without a protective net. Pumping speed for N2 is 1950 L/s with a protective net.
This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.