Standard Specifications of Pump
Model | TMP-B70 | |||
---|---|---|---|---|
Cooling method | Natural Air Cooling | Forced Cooling | Water Cooling | |
Ultimate pressure (Note 2) | after baking | (Note 1) | 10-8 Pa order (Note 2) | 10-8 Pa order (Note 2) |
before baking | 10-6 Pa
order |
10-6 Pa order | 10-6 Pa order | |
Pumping speed (Note 3) | N2 | 70 L/s | 70 L/s | 70 L/s |
He | 60 L/s | 60 L/s | 60 L/s | |
H2 | 49 L/s | 49 L/s | 49 L/s | |
Pumping speed(KF40) | N2 | 41 L/s | 41 L/s | 41 L/s |
He | 46 L/s | 46 L/s | 46 L/s | |
H2 | 42 L/s | 42 L/s | 42 L/s | |
Compression ratio | N2 | > 1×109 | > 1×109 | > 1×109 |
He | 4×107 | 4×107 | 4×107 | |
H2 | 4×105 | 4×105 | 4×105 | |
Critical Back Pressure (N2 gas、87,000rpm) | 1400Pa | 1400Pa | 1400Pa | |
N2 Gas Flow Rate for Continuous Operation (Note 4) | N2 | 300 Pa(temperature 25℃ or less) 200 Pa(temperature 30℃ or less) |
900 Pa(temperature 25℃ or less) 600 Pa(temperature 35℃ or less) |
900 Pa(temperature 25℃ or less) 600 Pa(temperature 40℃ or less) |
Maximum Outlet Pressure During Continuous Operation | 15 SCCM(temperature 25℃) 10 SCCM(temperature 30℃) |
40 SCCM(temperature 25℃) 30 SCCM(temperature 35℃) |
40 SCCM(temperature 25℃) 30 SCCM(temperature 40℃) |
|
Mass | ICF | 4.8 kg | 5.0 kg | 4.9 kg |
VG | 3.1 kg | 3.3 kg | 3.2 kg | |
ISO | 2.8 kg | 3.0 kg | 2.9 kg | |
KF | 2.9 kg | 3.1 kg | 3.0 kg | |
Bearing Type | Active Magnetic Bearing / Ceramic Ball Bearing | |||
Inlet flange | ICF114、VG65、ISO63K、KF40 | |||
Outlet flange | KF16 | |||
Rated spee | 87,000 rpm | |||
Start-up time(80 %まで) | 1.7 minutes | |||
Mounting position | All - direction | |||
Noise(by Shimadzu's method) (Note 5) | 48 dB(A)or less | |||
Admissible ambient magnetic field | Radial direction | 3 mT | ||
Axial direction | 15 mT | |||
Input electric power | Voltage | DC24 V ± 5 % | ||
Maximum power | 120 W |
※1: | Baking can only be performed under forced air cooling and water cooling specifications of the ICF flange. |
※2: | The values are for when a two-stage oil rotary pump (manufactured by Shimadzu Scientific Systems) is used as the auxiliary pump. |
※3: | Without a protective net. |
※4: | The values for continuous operation are for when the exhaust gas flow rate from the inlet is 0 mL/min. |
※5: | Measurement values for ISO flange specifications. |