Standard Specifications of Pump
Turbo molecular pump | TMP-2203LM(0) | TMP-2203LMC(0) | |
---|---|---|---|
Inlet flange | VG250 / ICF305 / ISO250B | ||
Outlet flange | KF40 | ||
Cooling method | Water | ||
Ultimate pressure (after baking) (Note 1) | 10-8 Pa order | 10-7 Pa order | |
Maximum allowable Argon gas flow rate | 1300 mL/min (Note 2) | 2000 mL/min | |
Maximum allowable inlet pressure (N2 continuous exhaust) | 40 Pa | ||
Maximum Outlet Pressure During Continuous Operation | 270 Pa | ||
Pumping speed (Note 3) | N2 | 2140 L/s | |
He | 2000 L/s | ||
H2 | 1350 L/s | ||
Compression ratio | N2 | 1.4×108 | |
He | 1×104 | ||
H2 | 1×103 | ||
Rated speed | 24600 rpm | ||
Start-up time | 11minutes | ||
Mounting position | In any desired direction | ||
Pump Inlet Baking Temperature | 120 ℃ or less | ||
Vibration level (by Shimadzu's method) | 0.01 μm or less(single amplitude) | ||
Recommended flow rate of purge gas | 20 ~ 30 mL/min (Note 2) | ||
Recommended pumping speed of backing pump in case of gas purge | 800 L/min or more | ||
Environment temperature | 0 ~ 40 ℃ | ||
Admissible ambient magnetic field | Radial direction | 3 mT | |
Axial direction | 15 mT | ||
Water | Flow Rate | 1 ~ 3 L/min | |
Pressure | 0.2 ~ 0.5 MPa | ||
Temperature | 5 ~ 30 ℃ | ||
Mass | 59 kg |
- (Note 1)The values are for when a metal gasket is used on the inlet flange.
- (Note 2)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
- (Note 3)Without a protective net. Pumping speed for N2 is 2050 L/s with a protective net.
Standard Specifications of Power Supply Unit
Power Supply | EI-R04M | |
---|---|---|
Battery | Unnecesssary | |
Combination Compatibility | Compatible combination of pump, power supply unit, magnetic bearing cable and fan cable are available. | |
Speed variation | Speed is variable between 25 % and 100 % of the rated speed. | |
Communication | Contact | REMOTE (MR-34) |
Input :START / STOP / RESET / LOW SPEED | ||
Output :ROTATION / ACC. / BRAKE / NORMAL / REMOTE / ALARM / WARNING | ||
Serial | Front panel:RS-232C (D-sub 9 pin male, Screw lock size M2.6) | |
Rear panel:RS-485 (D-sub 9 pin male, Screw lock size M2.6) | ||
Permissible Ambient Conditions | Temperature | During Operation 0 ~ 45 ℃ (No condensation is necessary condition. ) |
Humidity | 40 ~ 80 %RH | |
Input electric power | Voltage | Single phase AC 200 ~ 240 V ± 10 % (50/60 Hz ± 2Hz) |
Maximum power | 1.2 kVA | |
Mass | 8 kg |