Magnetic Bearings Turbo Molecular Pump, Integrated Power Supply TypeTMP-X3405

標準仕様

Turbo molecular pump TMP-X3405LMC TMP-X3405LMTC
Inlet flange VG300 / ISO320B / VG350
Outlet flange KF40
Cooling method Water
Ultimate pressure (after baking) 10-7 Pa order
Maximum allowable Argon gas flow rate 1300 mL/min (Note 1, Note 2) 800 mL/min(Set temperature:65 ℃) (Note 1, Note 4)
Maximum allowable inlet pressure (N2 continuous exhaust) 40 Pa
Maximum allowable outlet pressure
(N2 continuous exhaust)
270 Pa
Pumping speed (Note 3) N2 3400 L/s
Ar 3300 L/s
H2 2300 L/s
Compression ratio N2 2×108​
H2 1×103
Rated speed 27600 rpm
Start-up time 12 minutes or less
Mounting position In any desired direction
Vibration level
(by Shimadzu's method)
0.05 μm or less(single amplitude)
Set Temperature for Control - 65 ℃, 75 ℃ (Note 5)
Recommended flow rate of purge gas (Note 1) 30 mL/min 30 ~ 100 mL/min
Recommended pumping speed of backing pump in case of gas purge 2000 L/min or more
Environment temperature 10 ~ 40 ℃
Admissible ambient magnetic fie Radial direction 3 mT
Axial direction 15 mT
Water Flow Rate 3 L/min or more
Pressure 0.2 ~ 0.4 MPa
Temperatur 19 ~ 30 degrees C.(above dew point)
Displa LED POWER / ROTATION / ALARM POWER / ROTATION / TEMP.CONTROL / ALARM
Communication Contact REMOTE(D-sub 25 pin male, Screw lock size UNC 4-40)
Input:START / STOP / RESET / LOW SPEED
Output:ROTATION/ NORMAL / ACCELERATION / BRAKE / REMOTE / ALARM / WARNING Output:ROTATION/ NORMAL / ACCELERATION / BRAKE / REMOTE / TEMP.STABLE / ALARM / WARNING
Serial RS-232C / RS-485(Two-wire)
Speed variat Speed is variable between 25 % and 100 % of the rated speed. (set as 0.1 %)
Input electric power Voltage Single phase AC 200 ~ 240 V(50 / 60 Hz)
Maximum power 1.7 kVA 2.1 kVA
Mass 89 kg
  • (Note 1)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
  • (Note 2)The maximum allowable flow rate depends on the gas type, the APC valve temperature, the gas purge flow rate and outlet pressure.
  • (Note 3)Without a protective net. Pumping speed for N2 is 3000 L/s with a protective net.
  • (Note 4)The maximum allowable flow rate depends on the gas type, the APC valve temperature, set temperature of the pump, the gas purge flow rate and outlet pressure.
  • (Note 5)This is the temperature at which the cooling water is turned ON by the valve for temperature control in the pump section, using the monitored values from the temperature sensor. The heater will turn ON/OFF at the set temperature - 1°C.