Magnetic Bearings Turbo Molecular Pump, Integrated Power Supply TypeTMP-V4404

Standard specifications

Turbo molecular pump (Note 1) TMP-V4404LMW
Inlet flange VG350
Outlet flange KF40
Cooling method Water
Ultimate pressure (after baking) 10-7 Pa order
Maximum allowable Argon gas flow rate (Note 2, Note 3) 1450 mL/min
Maximum allowable inlet pressure
(Ncontinuous exhaust)(Note 3)
40 Pa
Maximum allowable outlet pressure
(N2 continuous exhaust)(Note 3)
170 Pa
Pumping speed (Note 4) N2 4050 L/s
Ar 4300 L/s
Compression ratio N2 1.5×107
He 4.0×103
H2 4.4×102
Rated speed 24000 rpm
Start-up time 12 minutes or less
Mounting position In any desired direction
Vibration level (by Shimadzu's method) 0.01 μm or less(single amplitude)
Recommended flow rate of purge gas (Note 2) 30 mL/min
Recommended pumping speed of backing pump in case of gas purge 1500 L/min or more
Environment temperature 10 ~ 40 ℃
Admissible ambient magnetic field Radial direction 3 mT
Axial direction 15 mT
Water (Note 5) Flow Rate 3 ~ 4 L/min
Pressure 0.2 ~ 0.4 MPa
Temperature 19 ~ 30 degrees C.(above dew point)
Display LED POWER / STATUS / ROTATION / NET
Communication Contact REMOTE(D-sub 25 pin male, Screw lock size #4-40UNC)
Input:START / STOP / RESET / LOW SPEED/ 
Output:ROTATION / ACC. / BRAKE / NORMAL /REMOTE / ALARM / WARNING / CONNECTION
Serial RS-232C / RS-485(D-sub 9 pin male, Screw lock size #4-40UNC)
Speed variation Speed is variable between 75 % and 100 % of the rated speed. (set as 0.1 %)
Input electric power Voltage (Note 6) Single phase AC 200 ~ 240 V(50 / 60 Hz)
Maximum power 1.2 kVA
Mass 98 kg
  • (Note 1)Do not use corrosive gas because there is possibility that the pump breaks down by the product material.
  • (Note 2)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
  • (Note 3)The maximum allowable flow rate depends on the gas type, the APC valve temperature, the gas purge flow rate and outlet pressure.
  • (Note 4)Without a protective net. Pumping speed for N2 is 2900 L/s with a protective net.
  • (Note 5)Water must keep temperature does not cause condensation.
  • (Note 6)Temporary voltage variation tolerance range: ±10%, temporary frequency variation tolerance range: 2 Hz.

This page may contain references to products that are not available in your country. Please contact us to check the availability of these products in your country.

Online Web Form

Inquiry