Magnetic Bearings Turbo Molecular Pump, Separate Power Supply TypeTMP-403

Standard Specifications of Pump

Turbo molecular pump TMP-403LM TMP-403LMC TMP-403M TMP-403MC
Inlet flange VG150 / ICF203 / ISO160B / ISO160C
Outlet flange KF25
Cooling method Water Air
Ultimate pressure (after baking) (Note 1) 10-8 Pa order 10-7 Pa order 10-8 Pa order 10-7 Pa order
Maximum allowable inlet pressure (N2 continuous exhaust) 200 Pa 1.3 Pa
Maximum Outlet Pressure During Continuous Operation 400 Pa 40 Pa
Pumping speed (Note 2) N2 420 L/s
He 360 L/s
H2 340 L/s
Compression ratio N2 1×109
He 8×104
H2 1×104
Rated speed 45000 rpm
Start-up time 5minutes or less
Mounting position In any desired direction
Pump Inlet Baking Temperature 120 ℃ or less
Vibration level (by Shimadzu's method) 0.012 μm or less(single amplitude)
Recommended flow rate of purge gas 20 ~ 30 mL/min (Note 3)
Recommended pumping speed of backing pump in case of gas purge 200 L/min or more
Environment temperature 0 ~ 40 ℃
Admissible ambient magnetic field Radial direction 3 mT
Axial direction 15 mT
Water Flow Rate 1 ~ 3 L/min
Pressure 0.2 ~ 0.5 MPa
Temperature 5 ~ 30 ℃
Mass 14 kg
  • (Note 1)The values are for when a metal gasket is used on the inlet flange. When an O-ring gasket is used, the achieved pressure is on the order of 10^-7 Pa.
  • (Note 2)Without a protective net. Pumping speed for N2 is 400 L/s with a protective net.
  • (Note 3)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).

Standard Specifications of Power Supply Unit

Power Supply EI-S04M
Battery Unnecesssary
Combination Compatibility Compatible combination of pump, power supply unit, magnetic bearing cable and fan cable are available.
Speed variation Speed is variable between 25 % and 100 % of the rated speed.
Communication Contact REMOTE(MR-34)
Input:START / STOP / RESET / LOW SPEED
Output:ROTATION / ACC. / BRAKE / NORMAL / REMOTE / ALARM / WARNING
Serial Front panel:RS-232C(D-sub 9 pin male, Screw lock size M2.6)
Rear panel:RS-485(D-sub 9 pin male, Screw lock size M2.6)
Permissible Ambient Conditions Temperature During Operation 0 ~ 45 ℃(No condensation is necessary condition. )
Humidity 40 ~ 80 %RH
Input electric power Voltage Single phase AC 100 ~ 120 V ± 10 %(50/60 Hz ± 2Hz)
Single phase AC 200 ~ 240 V ± 10 %(50/60 Hz ± 2Hz)
Maximum power 550 VA
Mass 8 kg

Online Web Form