Standard Specifications of Pump
Turbo molecular pump | TMP-803LM(0) | TMP-803LMC(0) | TMP-803M(0) | TMP-803MC(0) | |
---|---|---|---|---|---|
Inlet flange | VG150 / ICF203 / ISO160B | ||||
Outlet flange | KF40 | ||||
Cooling method | Water | Air | |||
Ultimate pressure (after baking) (Note 1) | 10-9 Pa order | 10-8 Pa order | 10-9 Pa order | 10-8 Pa order | |
Maximum allowable inlet pressure (N2 continuous exhaust) | 400 Pa | 1.3 Pa | |||
Maximum Outlet Pressure During Continuous Operation | 665 Pa | 40 Pa | |||
Pumping speed (Note 2) | N2 | 800 L/s | |||
He | 800 L/s | ||||
H2 | 700 L/s | ||||
Compression ratio | N2 | 1×109 or more | |||
He | 8×104 | ||||
H2 | 4×103 | ||||
Rated speed | 35000 rpm | ||||
Start-up time | 5minutes or less | ||||
Mounting position | In any desired direction | ||||
Pump Inlet Baking Temperature | 120 ℃ or less | ||||
Vibration level (by Shimadzu's method) | 0.01 μm or less(single amplitude) | ||||
Recommended flow rate of purge gas | 20 ~ 30 mL/min (Note 3) | ||||
Recommended pumping speed of backing pump in case of gas purge | 500 L/min or more | ||||
Environment temperature | 0 ~ 40 ℃ | ||||
Admissible ambient magnetic field | Radial direction | 3 mT | |||
Axial direction | 15 mT | ||||
Water | Flow Rate | 1 ~ 3 L/min | ― | ||
Pressure | 0.2 ~ 0.5 MPa | ||||
Temperature | 5 ~ 30 ℃ | ||||
Mass | 33 kg |
- (Note 1)The values are for when a metal gasket is used on the inlet flange. When an O-ring gasket is used, the achieved pressure is on the order of 10^-7 Pa.
- (Note 2)Without a protective net. Pumping speed for N2 is 730 L/s with a protective net.
- (Note 3)mL/min: volume flow rate at 0 degrees C., 1 atm (Compatible with SCCM).
Standard Specifications of Power Supply Unit
Power Supply |
EI-R04M
|
|
---|---|---|
Battery | Unnecesssary | |
Combination Compatibility | Compatible combination of pump, power supply unit, magnetic bearing cable and fan cable are available. | |
Speed variation | Speed is variable between 25 % and 100 % of the rated speed. | |
Communication | Contact | REMOTE(MR-34) |
Input :START / STOP / RESET / LOW SPEED | ||
Output :ROTATION / ACC. / BRAKE / NORMAL / REMOTE / ALARM / WARNING | ||
Serial | Front panel:RS-232C(D-sub 9 pin male, Screw lock size M2.6) | |
Rear panel:RS-485(D-sub 9 pin male, Screw lock size M2.6) | ||
Permissible Ambient Conditions | Temperature | During Operation 0 ~ 45 ℃ (No condensation is necessary condition. ) |
Humidity | 40 ~ 80 %RH | |
Input electric power | Voltage | Single phase AC 200 ~ 240 V ± 10 %(50/60 Hz ± 2Hz) |
Maximum power | 1.0 kVA | |
Mass | 8 kg |